Clinician Manual
Precision™ Spinal Cord Stimulator System Clinician Manual
en
Connecting to the IPG
For Percutaneous Lead - Dual Lead Connection
•
Superior (upper or left) leads, extensions, or 2x4 splitters connect to IPG port “1-L”.
•
Inferior (lower or right) leads, extensions, or 2x4 splitters connect to IPG port “2-R”.
For Infinion™ 16 Lead Connection
•
Laser-marked tail of Splitter 2x8 connects to IPG port “1-L”.
•
Unmarked tail of Splitter 2x8 connects to IPG port “2-R”
Lower
Upper
2R
1L
For Infinion CX Lead Connection
•
Proximal tail with one marker band connects to IPG port “1-L”.
•
Proximal tail with two marker bands connects to IPG port “2-R”.
For Percutaneous Lead - Single Lead Connection
•
Connect a single lead, extension, or 2x4 splitter to IPG port “1-L”.
•
Insert port plug into unused port.
Boston Scientific Confidential. Unauthorized use is prohibited.
Production
92070399 B.2
Precision Clinician Manual N EU Entrada
Page 54 of 394