Note: The 4 cm and 2.3 cm suture sleeves each have three (3) compression slots, which are
designed to reduce slippage.
4. Suture the sleeve to the supraspinous ligament or deep fascia through the suture sleeve
holes.
5. Tie multiple sutures as tightly as possible around the suture sleeve to secure it to the lead.
CAUTION: Tightening sutures directly on the lead can damage the lead.
6. For Permanent Trials, proceed to the instructions for “Tunneling the Lead or Lead
7. For Permanent IPG Implantation, proceed to the instructions for “IPG Implantation” on
Tunneling the Lead or Lead Extension
Note: If tunneling Infinion CX Lead(s), it is recommended to use the Long Tunneling Tool
(35 cm).
1. Attach the tunneling tool handle to the shaft by turning the locking mechanism clockwise.
Tool Handle
Locking Mechanism
Shaft