Operator’s Manual
7-2
Auxiliary Power Outlet Fuses
1.
Before replacing a fuse in the Controller, disconnect the mains power cord from the
Controller.
2.
Replace the fuse with another of the same type and rating. Refer to the fuse label on
the rear panel (see Figure 5-10).
3.
Use a slotted screwdriver to rotate the fuse holder counter-clockwise 1/8th of a turn.
The fuse holder “pops” out approximately 6 mm (1/4-inch).
4.
Remove the fuse holder and the fuse.
5.
Insert the new fuse in the fuse holder and reinsert it into the housing.
%
Call EPT for Service
If there appears to be a problem with the Controller,
please contact EPT for instructions on returning the Controller to EP