Controls and Displays
6-7
Controller in Power Control Mode - Catheter(T°) Connected
If the control mode is manually changed to Power Control, by pressing the RF POWER
CONTROL Button, the TEMPERATURE Display indicates 85
o
C during the READY
mode of the Controller, indicating the upper temperature limit at which the Controller
continues to deliver set power.
Controller in ON or CALIBRATE Operational Modes - Catheter(T°) Connected
When the Controller is in either the ON or CALIBRATE operational modes, the
TEMPERATURE Display indicates “LO” if the measured temperature is less than 20
°
C,
“HI” if temperature is greater than 120
°
C; otherwise, it shows the measured temperature
(see Figure 6-7).
Figure 6-7: Ready Mode, Initialization - Catheter(T°) Connected
Measured
Temperatur
Reading
( C)
TEMPERATURE
Catheter(Non-T) or No Catheter Connected
If a Catheter(Non-T) or no Catheter is connected, the display shows “---” (Figure 6-8).
Figure 6-8: Ready Mode, Initialization - Catheter(Non-T) Connected
Temperatur
Display
( C)
TEMPERATURE