Turning the RF Power Delivery “ON”
"
Important
Do not continue with this procedure unless all preceding procedures in this
chapter have been completed.
To deliver RF power to the Catheter, follow these steps:
1.
Ensure that all connections are secure and correct.
2.
Verify that all aforementioned selections have been made.
3.
To begin RF power delivery, press the RF POWER CONTROL Button once or hold
the Footswitch down. The POWER Display shows the RF power delivered to the
Catheter (in W).
The Audible tone
sounds during RF power delivery.
"
Note
If the Controller is in the Temperature Control mode and the measured
temperature is outside the range of 31
o
C to 43
o
C, the Controller does not deliver
RF power and an E07 error code message displays.
When RF power is delivered to the Catheter, the displays function as follows:
RF POWER CONTROL BUTTON/Light
This light illuminates and remains lit until RF power delivery is discontinued. (For
instructions on how to discontinue RF power delivery, see the section,
Turning the RF
Power Delivery “OFF”
on page 6-19.
POWER Display
This display shows (in W) RF power delivered to the Catheter. The value may fluctuate
slightly due to changes in tissue impedance.