Tachy Mode is programmed to:
–
Off
–
Monitor Only
–
M Therapy
Once you have programmed the pulse generator out of Storage mode, the device cannot be reprogrammed to
that mode.
X-RAY IDENTIFIER
The pulse generator has an identi
fi
er that is visible on x-ray
fi
lm or under
fl
uoroscopy. This identi
fi
er provides
noninvasive con
fi
rmation of the manufacturer and consists of the following:
•
The letters, BSC, to identify Boston Scienti
fi
c as the manufacturer
•
The number, 140, to identify the Model 2868 PRM software application needed to communicate with the
pulse generator
The x-ray identi
fi
er is embedded in the header of the device. For a left side pectoral implant, the identi
fi
er will
be visible by x-ray or
fl
uorography at the approximate location shown (Figure 1 on page 38).
37