Maintenance and Troubleshooting
Display
Description
Solution
E-6
Batteries are too weak for measurement.
Replace batteries with new ones, see chapter
“1.2 Activate your
Vivatmo me
”.
E-7
System self-test failed.
Replace batteries with new ones, see chapter
“1.2 Activate your
Vivatmo me
” and start again.
If the error persists, contact the service.
E-8
Regeneration failed.
Press the ENTER button to repeat the regeneration.
If the error persists, contact the service.
None – The
device will
not turn on.
Batteries are empty.
Replace batteries with new ones, see chapter
“1.2 Activate your
Vivatmo me
”.
Batteries are inserted the wrong way.
Ensure the batteries are inserted with correct polarity.
If you have any problems which you cannot solve with these Instructions for use or the device shows unexpected opera-
tions, contact your distributor or the
Vivatmo me
customer service center at [email protected]. For the contact data
see chapter “7 Manufacturer”. Please indicate the following information:
• Your name, address and phone number
• Serial number of the device (12-digit code on your
Vivatmo me
next to the
symbol at the bottom of the device)
• Error message
• A detailed description of the problem