© Bosch Healthcare Solutions GmbH
19
Instructions for use
Advanced Settings
en
4
Advanced Settings
WARNING
These settings shall be performed by healthcare professionals only. Incorrect operation may lead
to misinterpretation of the measuring results or deletion of the stored data.
If you have further questions, contact the
Vivatmo me
service center.
In order to access the advanced settings, press the ENTER button and the right ARROW button for
3 seconds.
4.1 Setting the upper and lower thresholds for the traffic light scale
The default values set for the traffic light scale are the reference values for adults in accordance with the ATS Guideline
2011, see chapter “2.4 Interpretation of results”.
1. Use the ARROW buttons to select the Thresholds icon. The icon starts flashing.
Press the ENTER button to select the flashing icon
2. Use the ARROW buttons to set the threshold value for the lower threshold. Confirm with ENTER
button.