Instructions for use
Maintenance and troubleshooting
6.2 Cleaning and disinfecting the analyser
Shut down
and
disconnect
the analyser from the power
supply.
To
clean
wipe the surface of the analyser with a c
leaning
tissue
.
For surface disinfection use a tissue soaked in
70%
ethanol/propanol-based
cleaning agent. If necessary,
use
DNAExitus
.
Only clean and disinfect the analyser on its exterior surfaces
and carefully wipe around openings.
• Clean the analyser including the scanner regularly.
• Disconnect the analyser from the power supply for cleaning.
6.3 Changing the air filter of the analyser
In order to ensure proper performance please make sure to replace the air filter of the analyser
every 6 months
. The
flap is part of the analyser. Make sure to close it after changing the air filter. Dispose of used air filters according to
regional and laboratory standards.
Shut down and turn the
analyser
off
and
unplug
all connections on the
back
.
Grasp the analyser with
both hands and
turn
it
over
its rearmost corner until it
is resting on its back.
Open
the flap on the analyser
bottom
and
remove
the used
air filter.
Insert
a new air filter
into the flap with the
soft surface side facing
towards the interior of the
analyser. Afterwards
close
the flap.
6.4 Quality control testing
It is strongly recommended to perform quality control testing regularly. For information about recommended
procedures please refer to the application-specific instructions for use.
6.5 Notice and error
Title of Notice
O K
Explanation
CTA and consequences. Optional: Contact data
I D :
X X X 0 - 9 9 C A
A
yellow dot
underneath the notice symbol indicates a new
entry.
Press
to open the list.
By selecting a notice further information will be displayed.
Read the explanation carefully.
4
C O N T I N U E
enabled
Press to go
back
and the notice
remains
in the list.
Press
ok
to close the notice and it will be
deleted
from the
list.
If an
error
is detected an error screen appears and the analyser is blocked. In case of error,
please
contact customer service
(for contact information see chapter 1.2).