Menu
5.1 Device settings
The greyed-out items are only accessible when logged in as an administrator.
4
C O N T I N U E
enabled
In the
menu
, select
settings
, then select
device settings
.
Select
display
and set the desired brightness by moving the slider.
Select
language
and choose the desired language from the list.
Select
date and time
to change the date and time of the analyser. Select “
>>
” to continue.
5.2 User accounts and passcode
The greyed-out items are only accessible when logged in as an administrator.
User 1
Administrator
User 2
Professional
User 3
Professional
User Accounts
PASSCODE ENABLED
A D D U S E R
4
C O N T I N U E
enabled
In the
menu
, select
settings
then select
user accounts
.
A list of registered users appears.
4
C O N T I N U E
enabled
Press
edit
to change name and/or passcode for the selected
account.
4
C O N T I N U E
enabled
4
C O N T I N U E
enabled
Activate
or
deactivate
an account.
4
C O N T I N U E
enabled
Press to
delete
an account.
4
C O N T I N U E
enabled
Press
add user
to create a new account. Select the
type
of account
:
• Administrator
• Professional
Afterwards type in a
username
and a
numeric passcode
.
4
C O N T I N U E
enabled
Enable or disable passcode
protection for professional
accounts.
• Only an administrator user can edit, activate/deactivate or delete the account of another
user. Professional users can only edit their own account.
• At least one administrator account must exist.
• When passcode protection is enabled, access is protected by a username and a numeric
passcode. A passcode is required to access an administrator account, even if passcode
protection is disabled.
• Deactivated accounts cannot be logged in until they have been activated again.
• The passcode consists of 8 digits for an administrator account and 4 digits for a professional
account.
5.3 Network configuration
A local area network can be configured to establish a connection to a HIS/LIS-System or Vivasuite.
The analyser supports two types of network connection:
• Ethernet
Make sure the
Ethernet cable
is connected to an
Ethernet port
at the back of the analyser.
• Wireless LAN (WLAN)
Log in as an
administrator
for access.
4
C O N T I N U E
enabled
In the
menu
, select
settings
, select
network/interfaces
, then select
network configuration
.