MIYN46 Wall Transmitter

Operation | en

21

Bosch Security Systems

User Manual US

970.035b | F.01U.278.521 | v2.0 | 2012.12

6.2

Legend of illustrations

6.3

Operation with the staff presence processing activated

6.3.1

Device behavior when pressing a button

State before action: Stand-by

 

** If the function Assistance only in case of Staff Presence is activated and the MIYN46 Wall 

Transmitter is not in Staff presence state, the Call for Assistance causes a state Alarm without 

Staff Presence and a Call for Help radio message.

Legend

Description

The contact of the Call for Help relay is closed

The contact of the Call for Help relay is open

The contact of the Call for Help relay opens and closes alternately

The contact of the Staff Presence relay is closed

The contact of the Staff Presence relay is open

The signal lamps are off

A signal lamp lights up in the indicated color

A signal lamp flashes in the indicated color

A signal lamp flashes quickly in the indicated color

Table 6.1

Action

State

Relay contact

Signal lamp

Radio Message

Stand-by

(none)

Call for Help

Alarm without staff 
presence

Call for Help

Call 
Acknowledgement

Staff presence 
without alarm

Acknowledgement

Call for Assistance

Assistance without 
staff presence **

Call for Assistance

Summary of Contents for MIYN46

Page 1: ...MIYN46 Wall Transmitter 970 035b F 01U 278 521 V2 0 2012 12 en User Manual US ...

Page 2: ......

Page 3: ...ption 8 3 1 1 Top view 8 3 2 Main functions 9 3 2 1 Red emergency call button 9 3 2 2 Blue assistance call button 9 3 2 3 Green acknowledgment button 9 3 2 4 Alarm signaling 9 4 Installation 10 4 1 Unpacking 10 4 1 1 List of contents 10 4 2 Installation 11 4 2 1 Generalities 11 4 2 2 Wall installation 11 4 2 3 Connecting to the power supply 12 4 2 4 Reset 12 4 2 5 Transmitter Identification 12 4 2...

Page 4: ...tion with the staff presence processing activated 21 6 3 1 Device behavior when pressing a button 21 6 4 Operation with the staff presence processing deactivated 24 6 4 1 Device behavior upon pressing a button 24 7 Storage 25 7 1 Short term storage conditions 25 7 2 Long term storage conditions 25 8 Disposal 26 8 1 Disassembly 26 8 2 Local disposal locations 26 8 3 Returning to the manufacturer 26...

Page 5: ...er allows the resident of an independent living or assisted living facility to easily alert the care personnel if needed The resident simply has to push the large red button The system is then able to control and to monitor the progress of the care personnel The alarm and event transmission is performed by radio There is no need for a wired connection between the MIYN46 Wall Transmitter and the Ma...

Page 6: ...s to property damages or to environmental pollution 2 2 Environmental conditions Install the MIYN46 Wall Transmitter in a dry place away from any source of heat WARNING The User Installer should read and understand this chapter before any intervention on the MIYN46 Wall Transmitter i NOTICE In case of unclear information please contact your local representative WARNING The MIYN46 Wall Transmitter ...

Page 7: ...ordance with the instructions and by authorized technical personnel only The sole possession of the User Manual US does not allow the personnel to perform any kind of repair on the MIYN46 Wall Transmitter Take into account all the warnings and follow all the instructions displayed on the MIYN46 Wall Trans mitter and those which are printed in the documentation Never try to use replacement pieces o...

Page 8: ...3 1 General description 3 1 1 Top view Fig 3 1 Top view 1 Red emergency call button 2 LED indicator for alarm 3 LED indicator for presence 4 Green acknowledgment button to acknowledge an alarm locally 5 Blue assistance call button 6 Socket for coded key insertion option 7 Installation compartment 8 Jack plug for pear push button ...

Page 9: ...press launching an emergency call 3 2 2 Blue assistance call button If upon arriving the staff requires additional help a call for assistance can be sent by pressing the round blue button 3 2 3 Green acknowledgment button In all cases there is an automatic call repetition until cancel of the call Cancelling can be made directly on the room transmitter with the smaller green button 3 2 4 Alarm sign...

Page 10: ...nsport 1 Take the components out of the box and place the MIYN46 Wall Transmitter in front of you 2 Check each component in the box in accordance with the list of contents below 3 Check that the MIYN46 Wall Transmitter and its accessories have not been damaged dur ing transportation 4 1 1 List of contents i NOTICE In case of defective or missing equipment do not try to install the MIYN46 Wall Tran...

Page 11: ...ws on a wall Two holes at 2 4 inches 60 mm interval are provided for direct mounting on connection boxes The three upper holes are used for positioning the device and cannot be reached from the installation compartment CAUTION Interferences Avoid immediate proximity to other electric devices such as a television 3 4 in 80 mm Ø 0 3 in 7 5 mm 2 2 in 53 8 mm 0 51 in 13 1 mm 0 15 in 3 7 mm 0 63 in 16 ...

Page 12: ...y make a short circuit on the two pins marked Reset before powering the device up again See Section Fig 4 2 Configuration and Wiring page 14 4 2 5 Transmitter Identification All MIYN46 Wall Transmitters are provided with their own radio identification code that is assigned in the factory During the installation it is necessary to set the relationship between the identification code of the Transmit...

Page 13: ...o not expose the battery to liquids Do not let the metal contacts on the battery touch another metal This could damage the battery Do not disassemble or modify the battery Do not expose the battery to extreme temperatures and never above 60 ºC 140 ºF For maximum battery capacity use the battery at room temperature Keep out of reach of children Use the battery for the intended purpose only Do not a...

Page 14: ...tance only by B7 B6 No processing B5 No Daily Message B4 Not used B3 Not used Staff Presence Call for Assistance from terminals J6 4 5 Staff Presence B1 REL 1 Call for Help B2 REL 2 Staff Presence o o polarity polarity N NC N NC NC NO NC NO J2 Polarity of Technical Alarm input J3 Polarity of input terminal J6 4 5 Connector J7 Presence Staff identification optional N NC J1 External Call for Help tr...

Page 15: ... causes the closing of the relay contact NC A Call for Help causes the opening of the relay contact Staff Presence Relay J5 5 6 B2 NO Staff Presence signaling causes the closing of the relay NC Staff Presence signaling causes the opening of the relay Call for Help J6 1 2 The contact closing causes a Call for Help Technical Alarm J6 3 4 J2 NO The contact closing causes a Technical Alarm NC The cont...

Page 16: ...dered Bridge State Function B3 Open Technical Alarms are not repeated B4 Soldered Calls are not repeated B5 Open A daily message is sent every 24 hours Soldered Daily messages are suppressed B6 Open o Staff Presence Processing is activated Soldered o Staff Presence Processing is disabled B7 Open Activation of J6 4 5 terminal input causes a Reserve Call Soldered Activation of J6 4 5 terminal input ...

Page 17: ...he Call for Help is repeated approximately every 2 minutes until a call acknowledgement takes place up to a maximum of 20 minutes This repeat function can be disabled by setting the solder bridge B4 A Call for Help is signaled by red flashes on the signal lamp and optionally by the activation of the relay Call for Help 5 2 Call for Assistance The Call for Assistance is triggered by pushing the Cal...

Page 18: ...ximum of 20 minutes This repeat function can be disabled by setting the solder bridge B4 The Reserve Call is signaled in the same way as the Call for Help on the signal lamp and the relays 5 5 Technical Alarm The Technical Alarm is triggered by activating the contact connected to terminals 3 and 4 of the terminal block J6 The alarm can be triggered by closing NO or opening NC the contact 3 Normall...

Page 19: ...ission can be checked by way of the daily message which is automatically sent about every 24 hours When the battery voltage is too low the daily message is replaced by a Battery Low message The duration of the daily message is 1 7 seconds The transmission of the Daily Message can be disabled by setting the solder bridge B5 this does not influence the transmission of the Battery Low message 5 9 Err...

Page 20: ...s on the device configuration of its actual condi tion and of the action completed as shown on the charts in the following chapters State Description Stand by No call is being processed Alarm without staff presence The resident has triggered a Call for Help but the Care Personnel has not yet responded Alarm with staff presence The Care Personnel has responded after a Call for Help Assistance witho...

Page 21: ... Description The contact of the Call for Help relay is closed The contact of the Call for Help relay is open The contact of the Call for Help relay opens and closes alternately The contact of the Staff Presence relay is closed The contact of the Staff Presence relay is open The signal lamps are off A signal lamp lights up in the indicated color A signal lamp flashes in the indicated color A signal...

Page 22: ...tate before action Alarm with staff presence Action State Relay contact Signal lamp Radio Message Alarm without staff presence none Call for Help Alarm without staff presence unchanged Call for Help Call Acknowledgement Alarm with staff presence Acknowledgement Call for Assistance Assistance without staff presence Call for Assistance Action State Relay contact Signal lamp Radio Message Alarm with ...

Page 23: ...staff presence Acknowledgement Call for Assistance Assistance with staff presence Call for Assistance Action State Relay contact Signal lamp Radio Message Assistance without staff presence none Call for Help Assistance without staff presence unchanged Call for Assistance Call Acknowledgement Alarm with staff presence Acknowledgement Call for Assistance Assistance without staff presence Call for As...

Page 24: ...l Acknowledgement Stand by Acknowledgement Call for Assistance Assistance without staff presence Call for Assistance Action State Relay contact Signal lamp Radio Message Alarm without staff presence none Call for Help Alarm without staff presence unchanged Call for Help Call Acknowledgement Stand by Acknowledgement Call for Assistance Assistance without staff presence Call for Assistance Action St...

Page 25: ...a longer time for instance in a storage room or any similar location it is recommended to use original packing material 7 1 Short term storage conditions Unwrapped MIYN46 Wall Transmitter in a room Protection against direct sunlight and dust 7 2 Long term storage conditions MIYN46 Wall Transmitter in its original packing material in a storage room Backup battery removed Protection against direct s...

Page 26: ...ocal waste collection ser vice or your local representative 8 1 Disassembly Only authorized personnel are allowed to disassemble a MIYN46 Wall Transmitter 8 2 Local disposal locations The nearest disposal locations are established in accordance with the currently applicable laws Ask the local authorities 8 3 Returning to the manufacturer If there is no practical disposal place the MIYN46 Wall Tran...

Page 27: ... to the following two conditions 1 This device may not cause harmful interference 2 This device must accept all interference received including interference that may cause undesired operation External power supply consumption 12 24 VDC Transmission 40 mA Idle mode 10 mA Battery power supply consumption 3V lithium battery CR2450 Transmission 30 mA Idle mode 10 µA Battery life Approx 2 years Table 9...

Page 28: ...Robert Bosch Healthcare Systems Inc 2400 Geng Road Suite 200 Palo Alto CA 94303 U S A Phone 650 690 9100 Toll Free 1 888 947 8957 Fax 650 798 3770 www bosch telehealth com ...

Reviews: