-26-
d
2,5 m
A
B
8 ft (2.5 m)
16 ft (5 m)
– The difference d of both marked points I and III on wall A indicates
the actual deviation of the measuring tool from the level plane.
The maximum permitted deviation dmax can be calculated as follows:
dmax = double the distance between the walls x 0.0118 in (0.3 mm)
Example: If the distance between the walls is 16 ft (5 m), the maxi-
mum deviation is dmax = 2 x 16 ft x 0.0118 in = 0.118 in (2 x 5m x
0.3 mm/m = 3mm). The marks must therefore be maximum 0.118 in
(3 mm) apart.
Checking the Leveling Accuracy of the Vertical Line
For this check, a door opening is required with at least 8 ft (2.5 m) of
space (on a firm surface) to each side of the door.
– Position the measuring tool on a firm, level surface (not on a tripod)
8 ft (2.5 m) away from the door opening. Allow the measuring tool to
level in while in cross-line operation mode, and direct the laser beams
at the door opening.