Start-up | 1 687 001 866 / 1 682 666 309 | 19
1 687 001 866 / 1 682 666 309 | 19
| 19
en
5.
Select "
Test functions >> Readjustment with
test gas
".
?
The "CO" dialog window is opened.
i
Numerical values must be entered separated by
a dot.
6.
Enter the sample gas value for CO and confirm
with <
Ok
>.
?
The "
CO2
" dialog window is opened.
7.
Enter the sample gas value for CO and confirm
with <
Ok
>.
?
The "
Propane
" dialog window is opened.
8.
Enter the sample gas value for propane and confirm
with <
Ok
>.
?
Zero position compensation is implemented.
The message
Set stable flow
is displayed.
Re-adjustment
1.
Connect the test hose to the calibrating gas inlet and
set a flow rate of ≥ 1 l/min.