![Boonton 4400A Manual Addendum Download Page 5](http://html.mh-extra.com/html/boonton/4400a/4400a_manual-addendum_2787853005.webp)
3.0 Corrections
1. Remove glitches that occur when in Triggered mode (as opposed to Auto) and certain
commands are executed. Also, measurement traces can now be moved and re-scaled
when Waiting for Trigger on the slow time bases.
2. Calculate the auto-measure parameter “OFF TIME”. This function has always been
enabled but there was no calculation method included.
3. Correct an overflow error in the cal table expansion that overwrites the first position
of the channel 2 table when channel 1 is expanded. This may cause the channel 2
PDF to not appear.
4. GPIB command TKBMEAS now reports the sign of Pk/Avg ratio correctly.
5. Marker math mode changes now occur immediately even in wait for trigger.
6. When both markers are in trigger view mode the marker math functions MK1-MK2,
MK2-MK1, MAX-MIN and MIN-MAX are computed as voltage difference and
appear in the middle window with voltage difference units. The PK/AVG mode is not
recognized in trigger view mode but is not an error. This correction also appears in
the parameters of GPIB commands TKMEAS, TKBMEAS and TKUNITS when
appropriate.
7. The trigger pointer is now removed when the direct set of a statistical mode occurs.
8. Restore the legacy GPIB command MKDELTA to set the marker math to power
difference in the linear units mode only. This command was deleted by mistake in the
“A” series.
9. Update the RUN/STOP message in the recall stored setup function to avoid out of
sync messages.
10. Change the linear mode reference lines to track vertical offset in “divisions” rather
than watts, which is incorrect.
Revised 20020513
98404706A