Boonton 4400A Manual Addendum Download Page 3

see below. Press the PLOT key to send the file to the selected output. The .bmp extension 
is added to the floppy disk file directory to allow viewing the filenames saved on disk.  
The GPIB commands for controller directed return of the .bmp file contents are: 
 

Send the sequence PLOTTER PLOT.BMP to select the bit-map mode. 
Send TKPLOT to set the talk mode that returns the file when addressed to talk. 

 
 The GPIB commands to send the plot file contents to an output other than the GPIB are: 
 

Send the sequence PLOTTER PLOT.BMP to select the bit-map mode. 
Send PLOTSER1, PLOTCOM1, PLOTLPT1 or PLOTDISK to select the output. 
{Use PLOT488 only with the front panel PLOT key manually to send in the talk 
only (ton) mode to a listen only (lon) device}. 
For PLOTDISK send FILENO ## to select a filename. 
Send PLOT to simulate pressing the PLOT key to send the file to the selected 
output. 

 
 

1.4 External Trigger Input Calibration 

 

A provision to zero and calibrate the external trigger inputs has been added to provide 
better accuracy for voltage measurements made with the trigger inputs. The following 
procedure is used to calibrate each external trigger input: 
 

Set 

Time>Timebase

 to “5 ms/Div” 

Set 

Trig>Trig Mode 

to “Auto” 

Select the external trigger input in the 

Trig>Trig Source

 menu corresponding to 

the selected measurement channel (CH1 to 1EXT or CH2 to 2EXT).  
Set the 

Chan#>Extensions>Display

 to “Trig” (Trigger View Mode) 

Set the 

Chan#>Vert Scale

 to 1.00 V/Div 

 Set 

the 

Spcl>Servicing>Cal Mode

 “On” 

 
With 

Cal Mode

 “On” two boxes labeled 

Ext Trig Zero

 and 

Ext Trig Cal

 will appear in the 

Chan#>Extensions

 menu. 

Ext Trig Zero

 will have a bright “Start” label.  

 
With no input to the selected external trigger input, press the menu button for 

Ext Trig 

Zero

 “Start”. The input will be zeroed and the 

Ext Trig Cal

 “Start” label will be bright.  

 
With +3.00 volts applied to the selected external trigger input, press the menu button for 

Ext Trig Cal

 “Start”. The input will be calibrated for 3 divisions of deflection at 1 V/Div.

 

 

  

 

Set the 

Spcl>Servicing>Cal Mode

 “Off” 

 

The results of the calibration are stored in non-volatile memory with file extension .TRV. 
Absent a file, default data is supplied automatically and simulates the existing software.

  

External trigger level calibration is not available on the GPIB. 

 

 

 

98404706A 

3

Summary of Contents for 4400A

Page 1: ...n and Off will be displayed The Off setting is volatile and not preserved through power cycles or major mode changes If the sensor in use does not have a temperature compensation table the Temp Comp menu box will not appear at all The Spcl Chan n Sensor report will include the message Sensor has Temperature Compensation Table when appropriate The Spcl Servicing Configuration report shows TC system...

Page 2: ...en with the new result If no previous file exists one is created Sensor filenames have the form SEN AC n where is the serial number and n is the channel number The file directory system is expanded to display auto calibration files in flash memory as well as the previous files on the floppy disk The Utility Disk Flash Disk path lists sensor auto cal files by channel The Select File menu contains a...

Page 3: ...s made with the trigger inputs The following procedure is used to calibrate each external trigger input Set Time Timebase to 5 ms Div Set Trig Trig Mode to Auto Select the external trigger input in the Trig Trig Source menu corresponding to the selected measurement channel CH1 to 1EXT or CH2 to 2EXT Set the Chan Extensions Display to Trig Trigger View Mode Set the Chan Vert Scale to 1 00 V Div Set...

Page 4: ...iles should be converted by reading the INS files and saving them as ISU The file directory is modified to display ISU files The GPIB commands are not affected 2 2 GPIB command OPT change An installed hardware options list has been appended to the OPT GPIB command format previously used For example a single channel instrument with Option 04 hardware installed and a sensor plugged in returns 1 1 0 ...

Page 5: ...er 6 When both markers are in trigger view mode the marker math functions MK1 MK2 MK2 MK1 MAX MIN and MIN MAX are computed as voltage difference and appear in the middle window with voltage difference units The PK AVG mode is not recognized in trigger view mode but is not an error This correction also appears in the parameters of GPIB commands TKMEAS TKBMEAS and TKUNITS when appropriate 7 The trig...

Reviews: