22
(Notice) If indicator is unmatched with load cell capacity or span
standard weight, indicator will display error message .
This value of span standard weight must be equal to full capacity,
or over 10% of full capacity.
( In case of less 1/5,000 resolution ,the value of standard weight
must be over 10% of full capacity at least.)
( In case of over 1/5,000 resolution ,the value of standard weight
must be over 20% of full capacity at least.)
(Notice) If span capacity is set less 10% , indicator will display
error message.
Key
Display
Description
Weight Load
Weight Value
Input (xxxxx)
Weight Load
Display
xx.xxx
xx.xxx
xx.xxx
xx.xxx
SPAn
xx.xxx
Load the weight which was
set in span value and press
“ENTER” key.
Under span calibration
Span calibration is
completed.
☞
REF
1. The weight shall be within the range of 10%~100% of maximum weight.
☞
REF
2. The setting weight must be over the range of 10% of maximum weight.
☞
REF
3. The setting weight over the maximum capacity.
☞ REF 4. If zero calibration is done without any error, “SPAn” message is
displayed the weight of setting weight is displayed on LED screen.
Check the weight.
☞
REF
5. Check the weight of setting weight.