▣ What is Simulation Calibration?
Calibration is performed for matching the BS-7300 to a strain-gage
sensor. The following two type of calibration are available for the BS-
7300.
%" G]ai`Uh]cb58>IGHA9BH
▣ Equivalent input Calibration.
Calibration is performed without an actual load by entering the rated
output value (mV/V) and the capacity (to be indicated) of the strain-
gage sensor by the keys. Calibration is easily performed when no actual
load is available.
For example, the gain is automatically determined by entering.
2.0000mV/V (rated output)-100.0kg (capacity)
as indicated for a load.
&" 7U`]VfUh]cbAYbi%GhYdr%$GhYd
1 Step : Calibration Mode Set
2 Step : Minimum Division Set
3 Step : Decimal Point Set
4 Step : Maximum Capacity Set &
Rated Output value Set
5 Step : Zero Calibration
6 Step : Digital Filter Value Set
7 Step : Relay mode Set
8 Step : Save
9 Step : END
GhYd%
- Function :
7U`]VfUh]cbAcXYGYh8]dgk]hW\gYh