Faults
General information
The following fault lists contain information
on faults which may occur on the
UVAHAND 250, possible causes and tips on
how to remedy the fault.
If a fault occurs on your equipment and can-
not be remedied by following these instruc-
tions, contact the customer service depart-
ment of the Bohle AG.
Contact address:
Bohle AG
Dieselstrasse 10, D-42781 Haan
Postbox 1163,
D-42755 Haan
Telefon: 0049 (0) 2129 5568-0
Telefax:
0049 (0) 2129 5568-201
Internet:
E-mail: [email protected]
Fault list Part 1
Fault
Cause of fault
Remedial action
Green lamp on
the mains
switch does not
light up
Device is not plugged in
to the mains.
•
Plug in to the mains, see: ‘Setting-
up, commissioning and operation;
Connecting up to the mains’.
No current at the mains
socket.
•
Check the main fuse.
Mains fuse is faulty.
•
Replace
fuse.
UV lamp does
not ignite
The ballast and the
UVAHAND 250 are not
connected.
•
Connect connecting cable, see:
‘Setting-up, commissioning and
operation; Connecting up to the
mains’.
The UVAHAND 250 is
cold. The ignition proc-
ess can take up to ap-
prox. 60 seconds.
•
Wait for ignition, see: ‘Setting-up,
commissioning and operation;
Switching the UVAHAND 250 on
and off’.
UV lamp is not fitted
correctly.
•
Fit UV lamp correctly, see: ‘Setting-
up, commissioning and operation;
Replacing the UV lamp’.
UV lamp is faulty.
•
Replace UV lamp, see: ‘Setting-up,
commissioning and operation; Re-
placing the UV lamp’.
UV lamp is still hot after
repeated use.
•
Allow the prescribed 3 to 5 minutes
for the equipment to cool off, see:
‘Setting-up, commissioning and
operation; Switching the
UVAHAND 250 on and off’.
Fault list Part 2
General
information
Fault list Part 1
Fault list Part 2