User’s Manual
25
Maintenance
Replacing the UV filter
Danger
Never operate the equipment if the filter is damaged or without a filter.
UV radiation is a health hazard!
The UV filter is a unit comprising a filter and
a filter frame.
The filter unit needs replacing if:
•
the filter is damaged.
•
a different radiation spectrum is required
for the application.
Warning
Use only spare parts manufactured by the supplier. The operational
safety of the UVAHAND 250 cannot be guaranteed if other parts are
used.
The following filter parts are available from
the Bohle AG:
Filter H1 with frame BO 5209406
Filter BL with frame BO 5209408
Note
To order spare parts, see: ‘Order data for equipment, spare parts and
accessories; Spare parts’.
The equipment must be opened in order to replace the UV filter.
Danger
Before opening the equipment, switch it off and disconnect from the
mains. There is a risk of serious, even fatal injury!
Let the lamp cool for at least 10 minutes before opening.
Acute danger of burns!
Maintenance
Replacing the
UV filter