8.2 

Replacement of Fuses 

 
There are two supply fuses located in the fuse drawer.  To change 
the fuses: 
 

Turn power switch to the off position 

 

Disconnect the unit from the power supply 

 

Remove the line cord from the power entry module on the 

 

back of the unit 

 

Pull back on the fuse drawer catch 

 

Pull out the fuse drawer (See Figure 3) 
Check and replace with the correct fuses if necessary.  The 
fuses should be 5mm x 20mm quick acting, rated 250V for 
230V units, 5mm x 20mm time delay, rated 250V for 115V units. 

 

 

Model 281111 (115 V AC):  

1.25AT 

 

 

Model 281111-2 (230 V AC): 

1.00AF 

 

Push the fuse drawer back in.  Reconnect unit to power 

 

supply. 

 
 

9.0  Warranty 

 
When  used  in  laboratory  conditions  and  according  to  these 
operating  instructions,  Boekel  warrants  this  product  to  be  free  of 
defective material and workmanship for a period of two years from 
the  date  of  manufacture.    The  liability  of  Boekel  for  any  defective 
equipment during the warranty period shall be limited to the repair 
of such equipment or replacement thereof without charge for parts 
or labor. 
 

Summary of Contents for Orbitron V 281111

Page 1: ...Boekel Orbitron V Models 281111 281111 2 Operating Instructions N2400331 Rev 2 01 2012...

Page 2: ...5 3 2 Installation 5 4 0 Operation 5 4 1 Identification of Controls See Figure 1 5 4 2 Timer Operation 6 4 3 Tray Angle Adjustment See Figure 2 6 5 0 Technical Specifications 7 6 0 Accessories 7 7 0...

Page 3: ...3 Figure 1 Rotator Front Panel Power Switch Speed Control Timer Control...

Page 4: ...ement Connect only to a power supply that provides a safety ground terminal Before moving disconnect at the power supply socket Do not remove the plug from the rear of the unit Ensure that the power s...

Page 5: ...n V Power line cord One 3 16 Hex key for tilt adjustment Operating instructions 3 2 Installation Place the Orbitron V on a clean flat and stable work surface to ensure that the suction feet on the bas...

Page 6: ...Remove load from Tray Rotate Speed Adjustment Knob counterclockwise until the Tray is rotating slowly When the pivot block is easily accessible for adjustment turn the Power Switch off Insert the supp...

Page 7: ...0 Hz Model 281111 2 230 VAC 10 0 7 A 50 60 Hz Power Rating Model 281111 50 W Model 281111 2 60 W Rotational Speed Variable from 5 to 120 rpm This product has been tested to the requirements of CAN CSA...

Page 8: ...ducts are designed to comply with IEC1010 1 No routine maintenance is required 8 1 Cleaning Disengage power cord prior to cleaning If a spill occurs use appropriate clean up procedures as required for...

Page 9: ...5mm x 20mm quick acting rated 250V for 230V units 5mm x 20mm time delay rated 250V for 115V units Model 281111 115 V AC 1 25AT Model 281111 2 230 V AC 1 00AF Push the fuse drawer back in Reconnect un...

Page 10: ...A Decontamination Certificate must be completed signed by the user and returned to Boekel Scientific prior to receiving the RMA number Please be sure to mark the outside of the returned goods package...

Reviews: