Boekel 133000 Operating Instructions Manual Download Page 7

 

7.     Technical Specifications 

 
This equipment is intended for indoor use and will meet its performance figures within 
the ambient temperature range of 10

°

C to 35

°

C, with maximum relative humidity of 80% 

(non-condensing).

  

Installation Category II (transient voltages).  Pollution Degree 2 in 

accordance with IEC 664.  Suitable for operation at altitudes of up to 6500 feet. 

 
Specifications: 

Temperature Range:   

 

(A5

°

C) to 60

°

Temperature Uniformity: 

 

+/- 1

°

C at 37

°

Temperature Display Resolution:  0.1

°

Supply Voltage Range: 

 

120V +/- 10%, 50/60 Hz 

   

 

 

 

 

230V +/- 10%, 50/60 Hz 

Power Rating:   

 

 

Model 133000:  90W 

                                                        Model 133000S:  90W 
   

 

 

 

 

Model 133000-2: 90W 

8.     Maintenance and Service 

All Boekel laboratory products are designed to comply with IEC1010-1.   No routine 
maintenance is required.  There are no user serviceable parts in this product.  
There are no parts that are required to be examined or supplied by Boekel or our 
agent. 

8.1 

Cleaning 

Disengage power cord prior to cleaning.  If a spill occurs, use appropriate clean 
up procedures as required for radiation or biohazard control.  The outer casing 
may be cleaned with water and a damp cloth.  Do not submerge or immerse the 
Incubator in water.  Before using any cleaning or decontamination method except 
those recommended by the manufacturer, users should check that the proposed 
method would not damage the equipment. 
 

8.2 

Replacement of Fuses 

There are two supply fuses located in the fuse drawer.  To change the fuses: 

 

Turn power switch to the off position 

 

Disconnect the unit from the power supply 

 

Remove the line cord from the power entry module on the back of the unit 

 

Pull back on the fuse drawer (see sketch below) 

 

Pull out the fuse drawer 

 

Check and replace with the correct fuses if necessary.  The fuses should be 
5mm x 20mm quick acting, rated 250V. 

Model 133000, 133000S (115 V AC):  

 

1AF 

             Model 133000-2 (230 V AC): 

 

.5AF 

 

Push the fuse drawer back in.  Reconnect unit to power supply. 

 

Summary of Contents for 133000

Page 1: ...Boekel Incubator Models 133000 133000 2 133000S Operating Instructions N2400181 Rev 1 05 11...

Page 2: ...r Lamps see Figure below 4 4 2 Setting the Temperature 5 4 3 Operating the Incubator 5 5 Accessories 5 5 1 Removable Shelf Catalog Number C1902053 5 5 2 Shelf Bracket Catalog Number A1903208 5 6 Fault...

Page 3: ...ety ground terminal Before moving disconnect at the power supply socket Do not remove the plug from the rear of the unit Do not check the temperature by touch but instead use the temperature display T...

Page 4: ...all shipping damage to the carrier immediately Shipping damage is covered by the carrier and repair replacement for shipping damages must be coordinated through the carrier Complete and return the Wa...

Page 5: ...er temperature in degrees Celsius The heater lamp indicated when the heater is operating The temperature adjust knob is used to control the chamber temperature Temperature Adjust Knob Power Switch O I...

Page 6: ...ff O position If the temperature adjust knob is not moved when the incubator is powered down the incubator will return to its previously set temperature upon restart 5 Accessories 5 1 Removable Shelf...

Page 7: ...xamined or supplied by Boekel or our agent 8 1 Cleaning Disengage power cord prior to cleaning If a spill occurs use appropriate clean up procedures as required for radiation or biohazard control The...

Page 8: ...ithout charge for parts or labor 10 Service It is required to obtain a Returned Material Authorization RMA number before any Boekel products are returned for any reason A Decontamination Certificate m...

Reviews: