4.2 

Setting the Temperature 

Use the temperature adjust knob to set the desired temperature.  The knob is 
rotated clockwise to increase the chamber temperature and counter clockwise to 
decrease the chamber temperature. 

4.3 

Operating the Incubator 

To turn the incubator on, switch the power switch to the on (I) position.  Set the 
desired chamber temperature.  The heater lamp will illuminate and remain lit until 
the chamber reaches the set temperature.  After reaching the set temperature, 
the heater lamp will cycle on and off with heater operation.  Allow a 60 minute 
chamber temperature equilibration when starting the incubator from a cold start.  
Allow at least 15 minutes for the re-equilibration when changing temperatures.  
To turn the incubator off, switch the power  switch to the off (O) position.  If the 
temperature adjust knob is not moved when the incubator is powered down, the 
incubator will return to its previously set temperature upon restart. 

5.     Accessories 

5.1 

Removable Shelf – Catalog Number C1902053 

5.2 

Shelf Bracket – Catalog Number A1903208 

6.     Fault Diagnosis 

 

Symptom 

Possible Cause 

Action Required 

1. Unit does not operate 

a.  Unit not switched on 
b.  Unit not plugged into 

power supply 

c.  Fuses blown 
d.  Power supply failure 

a.  Switch on 
b.  Plug in, switch on 
 
c.  Replace fuses per 8.2 
d.  Check that other 

electrical appliances on 
the same circuit are 
working 

 

2. Chamber temperature 

does not rise when 
expected 

a.  Actual temperature is 

higher than set 
temperature 

b.  Temperature control 

circuit fault 

a.  Check set temperature 
 
 
b.  Have unit checked by 

competent person 

 

3. Temperature continues 

to rise when not 
expected 

a.  Actual temperature is 

lower than set 
temperature 

b.  Temperature control 

circuit fault 

a.  Check set temperature 
 
 
Have unit checked by 
competent person 

Summary of Contents for 133000

Page 1: ...Boekel Incubator Models 133000 133000 2 133000S Operating Instructions N2400181 Rev 1 05 11...

Page 2: ...r Lamps see Figure below 4 4 2 Setting the Temperature 5 4 3 Operating the Incubator 5 5 Accessories 5 5 1 Removable Shelf Catalog Number C1902053 5 5 2 Shelf Bracket Catalog Number A1903208 5 6 Fault...

Page 3: ...ety ground terminal Before moving disconnect at the power supply socket Do not remove the plug from the rear of the unit Do not check the temperature by touch but instead use the temperature display T...

Page 4: ...all shipping damage to the carrier immediately Shipping damage is covered by the carrier and repair replacement for shipping damages must be coordinated through the carrier Complete and return the Wa...

Page 5: ...er temperature in degrees Celsius The heater lamp indicated when the heater is operating The temperature adjust knob is used to control the chamber temperature Temperature Adjust Knob Power Switch O I...

Page 6: ...ff O position If the temperature adjust knob is not moved when the incubator is powered down the incubator will return to its previously set temperature upon restart 5 Accessories 5 1 Removable Shelf...

Page 7: ...xamined or supplied by Boekel or our agent 8 1 Cleaning Disengage power cord prior to cleaning If a spill occurs use appropriate clean up procedures as required for radiation or biohazard control The...

Page 8: ...ithout charge for parts or labor 10 Service It is required to obtain a Returned Material Authorization RMA number before any Boekel products are returned for any reason A Decontamination Certificate m...