BOCK floorline 9.5 80 Assembly And Operation Manual Download Page 32

32 

 

4.5

 

Change of location 

If the bed must be moved to another location, please follow these safety instructions: 

-

 

Lower the lying surface with key (8) of the hand control. 

-

 

Before performing a  movement  with the nursing care bed  pull the mains plug 
and attach with the end piece to secure the power cable against falling and being 
crushed by anything. Make sure that the cable is not dragged over the floor. 

-

 

Before inserted the mains plug again, inspect the power cable visually for me-
chanical damage (dents and kinks, abrasions and bare wires). 

-

 

Place the power cable in a way that it will not be rolled over or strained during 
the operation of the bed or could be damaged when inserting the mains plug 
again. 

-

 

Perform the reset movement, as described on page 17. 

4.6

 

Transport, storage and operating conditions 

 

Transport and storage 

Operation 

Temperature 

0°C to +40°C 

10°C to +40°C 

Relative humidity 

20% to 80% 

20% to 70% 

Air pressure 

800hPa to 1060hPA 

4.7

 

Function notes 

To keep the bed in one location, lock the castor brakes on the chassis. To accomplish this, 
use your foot to move the locking lever on each castor downwards. Make sure that the 
castors are parallel to the longitudinal axis of the bed. This is important in the Trendelen-
burg and the anti-Trendelenburg position but also in the comfort sitting position. (Here, 
the brakes on one end piece must be unlocked.) 

If necessary, pull the integrated side rails up until they lock into place. When using mat-
tresses of different thickness, the minimum height of 22 cm, measured from the top edge 
of the side rail above the mattress without compression, may not be underrun (addition-
ally, a third side rail attachment guard must be used, which is available as an accessory). 

4.8

 

Disposal 

Each of the components made of plastic, metal and wood are recyclable and can be dis-
posed/recycled in compliance with the relevant legal provisions. Please note that electric 
adjustable nursing care beds or nursing beds are considered commercially used electronic 
scrap according to the WEEE-EC directive 2012/19/EC (b2b). All replaced  electrical and 
electronic components of the electrical adjustment  system must be handled in accord-
ance with the requirements of the Electrical and Electronic Equipment Act (ElektroG) and 
disposed of properly. 

Summary of Contents for floorline 9.5 80

Page 1: ...1 Assembly and operation manual Nursing care bed floorline 9 5 80...

Page 2: ...antee optimal lying comfort and allow professional care at the same time This product was designed with a focus on the elderly whose confi dence must be reinforced and whose life needs protection With...

Page 3: ...e hand control 18 3 7 Hand control lock functions 20 3 8 Second controller optional as accessory 20 3 9 Rechargeable Battery optional as accessory 21 4 Assembly and operation 23 4 1 Technical data 23...

Page 4: ...lled HomeCare bed 5 Home care nursing service 1 1 Intended purpose The nursing care bed is suitable for persons adults in need of care who are at least 146 cm tall The person s weight must not exceed...

Page 5: ...and cost units include all physical or juridical persons who use the beds or have the beds used for medical pur poses The briefing on the use of the product shall generally be conducted by the operat...

Page 6: ...f the person in need of care cannot operate the hand control safely due to cer tain disabilities if the person in need of care or the caregivers could be at risk due to those acci dental adjustments i...

Page 7: ...he manufacturer 1 5 Requirements for the installation location The company Hermann Bock GmbH is not liable for damages which might arise from the daily usage on the floor To avoid floor indentations f...

Page 8: ...and general type plate 1 Model designation 2 Manufacture date Day month and year 3 Serial number Order number running number 4 Mains voltage mains frequency and power input 5 Duty cycle 6 Drive protec...

Page 9: ...po sed of as household waste Medical application part type B Use only in dry rooms Protection class II double insulation insulated for protection IPX4 Protection of electrical equipment against splash...

Page 10: ...oltage DC motors 29 to 35V and con trolled with the smooth keys of the hand control ler The backrest can be adjusted electrically The leg part consists of a foot support that is divided into two parts...

Page 11: ...d panel at the gripping groove 4 and slowly raise the side rail somewhat Now you can fully press the release button 5 and slowly lower the side rail Please observe Be sure to raise the side rail sligh...

Page 12: ...oden side rail solo post in the middle Fig 3 Telescopic wooden side rail double post in the middle All dimensions in mm Depending on the length of the lying surface The single post at the head and leg...

Page 13: ...en side rail two bars 91703 Fig 2 Telescopic wooden side rail solo post in the middle Wooden side rail two bars 91868 Fig 3 Telescopic wooden side rail double post in the middle Wooden side rail long...

Page 14: ...s the thermal fuse immediately shuts off the power supply to the bed After a cooling time of approx one hour the power supply in the controller is automatically re stored 3 2 Caution Electric drive Th...

Page 15: ...p only to a minimum degree under load and it is equipped with electronic performance monitoring The result is a constant voltage up to the maximum load no loss of speed and a high level of protection...

Page 16: ...the controller sockets 3 4 5 and 6 The colour coding can be found in chapter 4 Assembly and operation A floor lighting can be connected to the connection socket for the hand control socket 2 which in...

Page 17: ...N BUS connection which can also be used to connect a sensor pad for a bed off detection Further details can be found in the separately supplied assembly and operation manual of the Smart Care Control...

Page 18: ...nance of the bed Thus a possible dis ruptive position of the hand control can be avoided by simply attaching it to any preferred spot on the bed Hazard note from Bock All drive components must not be...

Page 19: ...37 5 cm exit position safety stop at 27 5 cm safety position 9 Reset It is absolutely necessary to carry out this work during initial commissioning and after disconnection from the power supply The re...

Page 20: ...et 1 and the second controller in socket 2 are plugged into this second control box Sockets 3 4 and 5 remain unassigned and are equipped with a blind plug Hazard note from Bock Make sure that the cabl...

Page 21: ...d controller These functions must be locked separately here 3 9 Rechargeable Battery optional as accessory The battery serves as a mains independent reset function No warranty shall be provided for th...

Page 22: ...sively Then disconnect the battery from the controller and shut down the drive system Do not open or destroy the battery Do not expose the battery to heat or open fire Avoid storage in direct sunlight...

Page 23: ...nuous Wood side rail kg 165 Lying surface kg 66 Wood end panel incl lift column kg 39 Continuous wood side rail kg set 12 5 Telescopic wood side rail kg set 18 Electric data Manufacturer Dewert Baumei...

Page 24: ...tient s room or to allow access to the patient The nursing care bed is suitable for persons adults in need of care who are at least 146 cm tall The person s weight must not exceed 185 kg and must be o...

Page 25: ...lug the end piece with the narrow intake in the extension slide on the foot side of the lying surface connect the supplied hexagon socket screws M8x25 and tighten firmly 2 Screw the galvanized angle b...

Page 26: ...cable is also located Guide the cable of the lifting column from the foot part is through the still unused ca ble channel all the way to the controller Fix this cable at the angle bracket with a cable...

Page 27: ...them together After all side guard rails are installed slide the end piece back and tighten the screws again Telescopic side protection Place the outer posts at the end of the lying sur face and the m...

Page 28: ...table feet Move the bed to the lower position second safety stop and check whether there is still approx 5 mm of air left between the floor and the underside of the telescopic post Inserting the decor...

Page 29: ...ing the telescopic split side rail on this nursing care beds four adjustable feet must be screwed under the lying surface to prevent the telescopic posts of the side rail from touching the floor The s...

Page 30: ...ccessory and remove both attached aluminium caps Place the aluminium caps to the pull outs Push the foot end piece back again so that the longi tudinal frame and the part in position are aligned flush...

Page 31: ...e moves the locking pin from the motor The motor is now separated from the back part and can be swivelled downwards Once the second person has left the danger zone the first person can lower gently th...

Page 32: ...ccomplish this use your foot to move the locking lever on each castor downwards Make sure that the castors are parallel to the longitudinal axis of the bed This is important in the Trendelen burg and...

Page 33: ...stop after a short time There is an obstruction in the adjustment range Remove obstruction The safe working load has been exceeded Reduce the load The drives stop after a longer adjustment time The a...

Page 34: ...d of care who have a particular physique can only be achieved by means of custom built models With its customized models Hermann Bock GmbH enables customers to have their nursing care bed tailored to...

Page 35: ...itives for the cleaning of the plastic elements should generally be sufficient For care of the plastic surfaces use a product that is specifically suitable for plastics Drive To prevent the intrusion...

Page 36: ...excluded 2 Check all plug connections for correct position according to the instructions 3 Check the cables and electrical component parts for damage Should you detect any damage do not perform any c...

Page 37: ...Contact discharge 8 kV Air discharge 2 kV 4kV 8kV 15kV Floors should be wood concrete or ceramic tile If floors are covered with synthetic mate rial the relative humidity should be at least 30 Electro...

Page 38: ...e RF compliance level above the medizinisches Bett should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocat...

Page 39: ...rol and observance of the corresponding intervals for a fee The company Hermann Bock GmbH specifies an inspection interval which stipulates that a safety technical inspection is to be executed at leas...

Page 40: ...r damages to the wooden parts visible Electrical components 9 Power cables connecting cables and plugs without cable breaks pressure and kinking points abrasions porous points and exposed wires 10 Str...

Page 41: ...eater than 31 8 cm 34 Smooth running of the side rails in the tracks and safe locking Only nursing bed dino Smooth running of the doors on the aluminum profiles Secure locking of the doors into the lo...

Page 42: ...42...

Page 43: ...43...

Page 44: ...nd our sales partners can provide training supply of spare parts repairs inspections and other service Otherwise all warranty claims will be void A listing of our current distributors can be found und...

Reviews: