Emergency lowering - back part
In case of power or drive system failure, you
can lower the elevated back part manually.
Must be carried out always by two people!
One person lifts the back part slightly (to
take pressure off) and holds it in this posi-
tion. As next step, the second person re-
moves the locking pin from the motor.
The motor is now separated from the back
part and can be swivelled downwards.
Once the second person has left the danger
zone, the first person can lower gently the
back part.
Hold the back part by all means until it is
fully lowered.
Hazard note from Bock
Emergency lowering may be only carried out in an emergency by people who safely master this
operation.
Absolutely disconnect the bed from the mains as long as the motors have not been mounted
again.