To avoid body damage and prevent product and connected equipment damage,
carefully read the following safety information before using the test tool. This
product can only be used in the specified applications.
Warning:
To avoid fire or electrical shock, please use the proper power adapter. Use only the
power adapter appointed by the manufacturer and subject to the approval of being used
in the user's country.
Warning:
The two channels of the oscilloscope are non-isolated electrically. The channels should
adopt common basis during measuring. To prevent short circuits, the 2 probe ground
must not be connected to 2 different non-isolated DC level.
Warning:
The channels should adopt common basis during measuring. To prevent short circuits,
the 2 probe ground must not be connected to 2 different non-isolated DC level.
The diagram of the oscilloscope ground wire connection:
Ground Clip
Signal Input
Electrical Outlet
Probe
Oscilloscope
AC Adapter
The diagram of the ground wire connection when the oscilloscope powered by PC USB
is connected to the AC-powered PC through the USB ports:
Ground Clip
Signal Input
Oscilloscope
(PC USB power)
PC
Electrical Outlet
Probe
USB Cable
It is not allowed to measure AC power when the oscilloscope is AC powered through
the adapter, or when the oscilloscope powered by PC USB is connected to the
AC-powered PC through the ports.
Warning:
To avoid fire or electrical shock if a test tool input is connected to more 42V peak
(30Vrms) or on circuits of more than 4800VA:
●
Use only insulated probes and adapter supplied with the test tool, or indicated by
BNC as suitable for the Oscilloscope.
●
Before use, inspect probes and accessories for mechanical damage and replace
when damaged.
●
Remove all probes and accessories that are not in use.