BNC P4017 Quick Start Manual Download Page 6

To avoid body damage and prevent product and connected equipment  damage, 
carefully read the following safety information before using the test tool. This 
product can only be used in the specified applications.   

Warning:

 

To avoid fire or electrical shock, please use the  proper  power  adapter. Use only the 

power adapter appointed by the manufacturer and subject to the approval of being used 

in the user's country.   

Warning: 

The two channels of the oscilloscope are non-isolated electrically. The channels should 

adopt common basis during measuring. To prevent short circuits, the 2 probe ground 

must not be connected to 2 different non-isolated DC level.   

Warning: 

The channels should adopt common basis during measuring. To prevent short circuits, 

the 2 probe ground must not be connected to 2 different non-isolated DC level.   

The diagram of the oscilloscope ground wire connection:   

Ground Clip

Signal Input

Electrical Outlet

Probe

Oscilloscope

AC Adapter

The diagram of the ground wire connection when the oscilloscope powered by PC USB 

is connected to the AC-powered PC through the USB ports:   

Ground Clip

Signal Input

Oscilloscope

(PC USB power)

PC

Electrical Outlet

Probe

USB Cable

It is not allowed to measure AC power when the oscilloscope is AC powered through 

the adapter, or when the oscilloscope powered by PC USB is connected  to the 

AC-powered PC through the ports.   

Warning: 

To avoid fire or electrical shock if a test tool input is connected to more 42V peak 

(30Vrms) or on circuits of more than 4800VA: 

Use only insulated probes and adapter supplied with the test tool, or indicated by

BNC as suitable for the Oscilloscope.

Before use, inspect probes and accessories for mechanical damage and replace

when damaged.

Remove all probes and accessories that are not in use.

Berkeley Nucleonics Corporation  

[email protected] | www.berkeleynucleonics.com 

2955 Kerner Blvd, San Rafael, CA 94901 | 800-234-7858

Summary of Contents for P4017

Page 1: ...Model P4017 Single Channel USB Oscilloscope Quick Start Guide...

Page 2: ...f customer proof of purchase This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care BNC shall not be obligated to furnish s...

Page 3: ...spection 6 5 How to Connect with PC 6 6 How to Implement the Function Inspection 7 7 How to Implement the Probe Compensation 7 8 How to Use the Probe Safely 8 9 Appendix 9 Appendix A Enclosure 9 Appen...

Page 4: ...allowed on bare conductors When the product is powered on do not touch any bare joints or parts of the scope meter The operation is prohibited in case of any undetermined failure When in doubt any da...

Page 5: ...rms may appear on this product Danger It indicates an injury or hazard may immediately happen Warning It indicates an injury or hazard may be accessible potentially Caution It indicates potential dama...

Page 6: ...round wire connection Ground Clip Signal Input Electrical Outlet Probe Oscilloscope AC Adapter The diagram of the ground wire connection when the oscilloscope powered by PC USB is connected to the AC...

Page 7: ...ly transmitted to the Oscilloscope Do not touch the exposed part of metal BNC Do not insert metal objects into connectors Always use the Oscilloscope only in the manner specified Voltage ratings that...

Page 8: ...ptional Network port which can be used to connect with PC 5 USB port Supply power by PC USB or the adapter communicate with PC 6 Signal input of Channel 4 7 Signal input of Channel 3 8 Signal input of...

Page 9: ...mally or fails in the performance test please get in touch with the BNC s distributor responsible for this business or the BNC s local offices If there is damage to the instrument caused by the transp...

Page 10: ...to Implement the Probe Compensation When connecting the probe with any input channel for the first time make this adjustment to match the probe with the input channel The probe which is not compensate...

Page 11: ...inger against any electric shock shown as Figure 8 1 Figure 8 1 Finger Guard Warning To avoid electric shock always keep your finger behind the safety guard ring of the probe during the operation To p...

Page 12: ...ollowing steps 1 Wipe the dust from the instrument and probe surface with a soft cloth 2 Disconnect the USB cable before cleaning your Oscilloscope Clean the instrument with a soft wet cloth not dripp...

Reviews: