6. How to Implement the Function Inspection   

Make a fast function check to verify the normal operation of the instrument. 
First, connect the oscilloscope with PC successfully, and then do the following 
steps: 

 

1. 

Click 

  in the software interface to restore to factory settings. The default 

attenuation coefficient set value of the probe in the software is 10X. 

2. 

Set the Switch in the Oscilloscope Probe as 10X and Connect the 

Oscilloscope with CH1 Channel.

   

Align the slot in the probe with the plug in the CH1 connector BNC, and then 
tighten the probe with rotating it to the right side.   

Connect the probe tip and the ground clamp to the connector of the probe 
compensator. 

3. 

Click 

  on the right top of the software interface to auto set.   

The square wave of 1 kHz  frequency and 3.3  V peak-peak value  will be 
displayed. 

 

4. 

Check CH2 by repeating Step 2 and Step 3.   

(P4017 needs to continue to check CH3 and CH4). 

7. How to Implement the Probe Compensation   

When connecting the probe with any input channel for the first time, make this 
adjustment to match the probe with the input channel. The probe which is not 
compensated or presents a compensation deviation will  result in the 
measuring error or mistake. For adjusting the probe compensation, please 
carry out the following steps:   

1.  Click 

  in the software interface  to show the function menu, choose 

"

Channel

". Click to select "

CH1

", set the "

Probe Rate

" in the menu as 

10

Set  the switch in the probe as 10X

 

and connect the probe with the CH1 

channel. If a probe hook tip is used, ensure that it keeps in close touch with 

the probe. Connect the probe tip with the signal connector of the probe 

compensator and connect the reference wire clamp with the ground wire 

connector of the probe connector, and then click 

  on the right top of the 

software interface to auto set.   

 

The set values of the probe switch are 1X and 10X 

Figure 7-1 Attenuation Switch   

Berkeley Nucleonics Corporation  

[email protected] | www.berkeleynucleonics.com 

2955 Kerner Blvd, San Rafael, CA 94901 | 800-234-7858

Summary of Contents for P4017

Page 1: ...Model P4017 Single Channel USB Oscilloscope Quick Start Guide...

Page 2: ...f customer proof of purchase This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care BNC shall not be obligated to furnish s...

Page 3: ...spection 6 5 How to Connect with PC 6 6 How to Implement the Function Inspection 7 7 How to Implement the Probe Compensation 7 8 How to Use the Probe Safely 8 9 Appendix 9 Appendix A Enclosure 9 Appen...

Page 4: ...allowed on bare conductors When the product is powered on do not touch any bare joints or parts of the scope meter The operation is prohibited in case of any undetermined failure When in doubt any da...

Page 5: ...rms may appear on this product Danger It indicates an injury or hazard may immediately happen Warning It indicates an injury or hazard may be accessible potentially Caution It indicates potential dama...

Page 6: ...round wire connection Ground Clip Signal Input Electrical Outlet Probe Oscilloscope AC Adapter The diagram of the ground wire connection when the oscilloscope powered by PC USB is connected to the AC...

Page 7: ...ly transmitted to the Oscilloscope Do not touch the exposed part of metal BNC Do not insert metal objects into connectors Always use the Oscilloscope only in the manner specified Voltage ratings that...

Page 8: ...ptional Network port which can be used to connect with PC 5 USB port Supply power by PC USB or the adapter communicate with PC 6 Signal input of Channel 4 7 Signal input of Channel 3 8 Signal input of...

Page 9: ...mally or fails in the performance test please get in touch with the BNC s distributor responsible for this business or the BNC s local offices If there is damage to the instrument caused by the transp...

Page 10: ...to Implement the Probe Compensation When connecting the probe with any input channel for the first time make this adjustment to match the probe with the input channel The probe which is not compensate...

Page 11: ...inger against any electric shock shown as Figure 8 1 Figure 8 1 Finger Guard Warning To avoid electric shock always keep your finger behind the safety guard ring of the probe during the operation To p...

Page 12: ...ollowing steps 1 Wipe the dust from the instrument and probe surface with a soft cloth 2 Disconnect the USB cable before cleaning your Oscilloscope Clean the instrument with a soft wet cloth not dripp...

Reviews: