background image

BMG LABTECH 

    

FLUOstar / POLARstar / LUMIstar OPTIMA Operating Manual 

2007-11-19 0413B0001I 

21/23 

 

4.6 Reagent 

Injectors 

The FLUOstar, POLARstar as well as the LUMIstar OPTIMA can all be equipped with up to 2 reagent 
injectors (figure 33). 
 

 

 
 

 

 

 

       Figure 33: Reagent injectors 

 

When the reagent injector(s) are not in use, the needle(s) can be placed in the needle holder (figure 8). 

The reagent needles are made of stainless steel, the tubings and valve housing are made of Teflon and Kel-
F, and the syringe barrel is made of glass. All materials are among the most chemical resistant materials 
available. 

Of course, the needle tip plays a role regarding the pumps' accuracy. You should always treat the needles 
with care. That is, e.g. be carefull when positioning the needles in the measurement head or in the needle 
holder. 

For obtaining optimal performance of the reagent injectors, please see the following chapter. 

4.6.1  Use and Maintenance of the Reagent Injectors 

To remove cellular debris and viscous solutions from the syringe barrel: 
Take off the syringe barrel and rinse it with distilled water. It may be useful to use the wire syringe cleaners 
(can be found in the service box) to scrape particles off the walls. 

In order to obtain optimal performance from the reagent injectors, it is recommended to follow these 
guidelines in the use of the reagent injectors: 

 

Do not use the syringes more than two cycles without liquid. 

 

After each use, thoroughly flush the syringes with distilled water. 

 

If the plunger is removed from the syringe barrel, it should be wiped with ethanol before 
replacing. 

 

Syringes should be cleaned each week using one of the following procedures: 

 

Cleaning with weak detergent or 10% bleach 

1.  Fill the syringe with a weak detergent or 10% bleach solution 
2.  Leave the solution in the syringe for 30 minutes 
3.  Flush the syringe a minimum of 10 times with distilled water 

 

Cleaning with acid / base 

(best procedure if cells are used in the syringe)

 

1.  Fill the syringe with 0.1M NaOH and leave it in the syringe for 10 minutes. 
2.  Flush the syringe with distilled water. 
3.  Fill the syringe with 0.1M HCl, and leave the solution in the syringe for 10 minutes. 
4.  Flush the syringe a minimum of 10 times with distilled water. 

Summary of Contents for FLUOstar OPTIMA

Page 1: ...Operating Manual Revision I...

Page 2: ...3930 94500 Champigny s Marne Omiya ku 330 0844 Phone 61 3 5973 4744 Phone 33 1 48 86 20 20 Saitama City Japan Fax 61 3 5973 4711 Fax 33 1 48 86 47 07 Phone 81 48 647 7217 australia bmglabtech com fra...

Page 3: ...ure in the above mentioned modes and additionally in fluorescence polarization mode and simultaneous dual emission The LUMIstar OPTIMA is a luminescence microplate reader that can be upgraded to inclu...

Page 4: ...12 4 2 2 FLUORESCENCE INTENSITY OPTICS 13 4 2 3 ABSORBANCE OPTICS 13 4 2 4 LUMINESCENCE OPTICS 14 4 2 5 FLUORESCENCE POLARIZATION OPTICS 14 4 2 6 DUAL EMISSION OPTICS 15 4 2 7 HIGH DENSITY OPTICS 15 4...

Page 5: ...filter positions Gain control Software selectable gain Automatic gain adjustment Plate Carrier Auto lock microplate carrier All microplate formats up to 1536 well in all detection modes Microplates sh...

Page 6: ...10 OD for 0 2 OD range 0 030 OD for 2 3 OD range Computer interface RS232 38400 baud binary communication protocol Power requirements 100 115 230 V 50 60 Hz Consumption max 205 VA Fuses T2 5A 250V for...

Page 7: ...ng and positioning the reader make sure to unlock the transport pin section 2 1 Transport Pin before any power connection section 2 2 Power and Communication Connections 2 1 Transport Pin When the ins...

Page 8: ...ment s power cable and turning the power switch on make sure the transport pin is unlocked section 2 1 Power Connection First check that the power switch on the back of the instrument is in the Off po...

Page 9: ...A flashing green light means the instrument is busy e g performing a measurement plate in out priming etc A faster flashing 5 flashes per second means an error has occurred Back View On Off Power cabl...

Page 10: ...8 FLUOstar OPTIMA POLARstar OPTIMA top view of reagent box with combi optic and 2 reagent injectors Top View Reagent Box LUMIstar OPTIMA Transport pin Figure 9 LUMIstar OPTIMA top view of reagent box...

Page 11: ...ion and Changing of Optics The readers are equipped with quick fix mountings for easy exchange of optics and easy to position spacers for more about spacers see 4 5 No tools are required for changing...

Page 12: ...ging of Optics For fluorescence measurements excitation enters through the yellow marked light guide and emission is measured through the blue marked light guide For absorbance measurements The grey r...

Page 13: ...tics Regarding reagent injection be careful when you position the needles in the measurement head see figure 15 to avoid damage to the reagent needles as well as to the optic The optional reagent inje...

Page 14: ...Luminescence optics Regarding reagent injection be careful when you position the needles in the measurement head see figure 18 to avoid damage to the reagent needles as well as to the optic The option...

Page 15: ...ptics look like the polarization optics but are not capable of polarization The optic has to be installed in the same way as the polarization optic figure 20 4 2 7 High Density Optics The high density...

Page 16: ...Well Injection Head A special injection head is required in order to inject into 384 well plates The injection head is installed similarly to an optic see section 4 2 Installation and Changing of Opti...

Page 17: ...installation of the software the filter must be typed in the filter table figure 25 In the Control software the filter table can be reached by choosing Setup Filter see the software manual for additio...

Page 18: ...eful not to touch any filters and turn the nut counter clockwise The filter wheel can be pulled straight out of the housing The filter positions are numbered 1 to 8 When installing new filters note th...

Page 19: ...Fluorescence Polarization Filters The excitation filters can go in any available position in the excitation wheel The emission wheel has two filters of the same wavelength for each channel They shoul...

Page 20: ...ier see figure 31 spacers needs to be installed under the measurement head see figure 32 There should be enough spacers so that the height of the left side of the plate carrier is slightly higher than...

Page 21: ...nge barrel Take off the syringe barrel and rinse it with distilled water It may be useful to use the wire syringe cleaners can be found in the service box to scrape particles off the walls In order to...

Page 22: ...disinfecting the instrument use the following solutions Alcohol 70 Authorized personnel wearing disposable gloves and protective clothing should only perform the disinfection procedure The location sh...

Page 23: ...disinfected according to the instructions of the operating manual of instrument Name ________________________________________________ Company ________________________________________________ ________...

Reviews: