background image

meM-LOG-SE 

 

 

© BMC Messsysteme GmbH 

 

Page 7 

5  Using the BP16 

If not only voltages are measured but also 
other physical quantities like temperature, 
acceleration, resistance etc. using sensors, 
measuring amplifiers are necessary for signal 
conditioning. 

Especially comfortable is the connection of a 

BP16

 to the data logger, since the carrier 

board for measuring amplifiers is mounted 
into the same housing. If you fit the devices at 
the red frames together, they form a stable 
unit. The connection of a BP16 to the 

meM-

LOG-SE

 is directly possible via a 1:1 connec-

tion of the two 37-pole Sub-D female connec-
tors with a 37-pole ribbon cable. 

The 5V auxiliary voltage, led out at pin 17 of the 37-pole Sub-D socket of the logger, supplies the 

BP16

 with power 

and turns it on and off. 

6  Operating software ST-meM-LOG SE 

With the configuration software 

ST-

meM-LOG

 

SE

, you can easily test the 

functions of the logger via USB or 
RS232 interface, configurate the device, 
start scans or read out measuring data.  

For further help, please have a look at 
the respective user manual, which is 
also included, or the integrated online 
help. When developing the second 
version, (SE: "Second Edition") great 
emphasis was put on user-friendliness 
and reliability. 

 

 

 

¾

 

Administration of USB and the serial interface in one program 

The connection between PC and data logger is realized via the serial (RS232) as well as the USB interface so that 
the advantages of both interfaces are available to the user (see chapter 3.3). ST-meM-LOG SE manages both inter-
faces so that one device (or several) can be connected to USB and another to the serial interface at the same time.  

¾

 

Connection of several devices 

Using the USB interface, up to 127 devices can be operated with ST-meM-LOG SE. To connect more than one log-
ger, the serial interface RS485 must be used. This requires a converter between RS232 and RS485. In this case 255 
devices at the most can be used.  

¾

 

Live display 

Incoming live data of the analog and digital inputs of the selected device are continuously displayed as numerical 
values or digital levels (0 = 

low

, 1 = 

high

). The 16 analog channels are presented as physical quantities within their 

adjusted measuring range with physical unit and channel name.  

N

N

E

E

W

W

 

 

!

!

 

 Analog input values are displayed as physical quantities if adjusted and not as simple voltages. Saved measur-

ing values are shown in the live display during a scan in 

Continuous Mode

 or in 

Alarm Mode

 (saving all chan-

nels in alarm). Channel values that are not scanned are removed from the display. Alarm states are signalized 
by a red marking. 

Summary of Contents for MEM-LOG-SE

Page 1: ... it is no problem to realize measuring applications which require mobility and flexibility in a high degree In order to extend the operating time of the accumu lator the device changes into an energy saving mode after 1 minute Sleep Mode and just wakes up occasionally e g shortly before scanning the next measuring value Via 16 analog input channels in the measuring range of 5V with a resolution of...

Page 2: ...ition and processing measuring files created with ST meM LOG SE can be dis played in NextView 4 as graphic signal curves and then be ana lyzed NextView 4 is also available as a stripped down Analysis version A free demo of the software is included with delivery The instruction set for program ming the meM LOG SE is com patible to the ADAM4000 series For further information please visit our homepag...

Page 3: ... each channel separately storage of either all channels to be saved or only of channels which are in state of alarm If measuring values are within their normal range again the scan is temporarily interrupted If not in Loop Mode see below the scan continues until memory is full or measurement is stopped even if a scan has been stopped with the device button it can be taken up again the new measurin...

Page 4: ...listed in the following table Pin Sub D15 meM LOG SE 1 Alarm Out H active high 9 Alarm Out L active low 8 15 Relay contact alarm 3 Digital In 1 active low 11 Digital In 2 active low 12 Scan start active low 5 13 Ground 2 4 6 7 10 14 n c In case of alarm the alarm outputs and the relay contact at pin 1 9 or 8 15 become active at the same time as long as the signal values exceed the defined range Al...

Page 5: ...ion over great distances connection of a modem 3 3 1 Serial connection RS232 The serial connection requires a 9 pole cross wired null modem cable not electrically isolated which is connected to the two 9 pole Sub D plugs of the logger and the PC to be ordered under ZUKA SER9 The serial connection of the logger is located at the back of the device If your PC only provides a 25 pole serial connectio...

Page 6: ...scan Æ turn off device measuring values are saved up until the point of turning off at least 5sec Scan running repeated and short flashing of LED fast flickering after 0 5sec after 5sec LED off Instant On temporarily inter rupt energy saving mode 0 5sec Sleep Mode not changing If you start a scan with the device button or a digital signal at the digital connector the memory must be empty This does...

Page 7: ... was put on user friendliness and reliability Administration of USB and the serial interface in one program The connection between PC and data logger is realized via the serial RS232 as well as the USB interface so that the advantages of both interfaces are available to the user see chapter 3 3 ST meM LOG SE manages both inter faces so that one device or several can be connected to USB and another...

Page 8: ... possible in all operating modes Store measuring data with different file format When reading out measuring data from the memory of the data logger ST meM LOG SE writes files in txt or lfx file format which can be displayed as table values and analyzed in various programs for word processing Notepad Wordpad Word etc and spreadsheet e g Excel Especially to mention is the storage as an lfx file the ...

Page 9: ... the software automatically SP2 do not connect with Windows Update Windows 2000 Search for a suitable driver for my device Under Windows Vista no selections have to be made The driver is found and installed automatically upon connection of the hardware because the driver package has been installed on hard disc before For the same reason no additional location needs to be entered for the driver sea...

Page 10: ...0Hz channel Sampling period 18h 12min 15sec 1sec in steps of sec 10min 55sec 350msec 0 01sec in steps of 1 100 sec Total sampling rate 16Hz 1 6kHz The total sampling rate is the sum of the sampling rates of the individual used channels e g 5 channels à 100Hz 500Hz total sam pling rate The values for accuracy always relate to the respective measuring range Errors might add at worst Digital inputs a...

Reviews: