Pin 12 features all functions of the device button (see chapter 4). That means if you connect pin 12 to ground, a scan 
will be started. In this case, the pin must be short-circuited for more than 0.5sec and less than 4sec. 

 

 

The digital input and output channels are protected with 1k

 resistors. If the input voltage is not within the 

admitted voltage range of 0V..5V, the device may be damaged. 

 

The signals of digital input channels are "hidden" among the analog measuring values, so that always at 
least one analog input channel must be saved! 

 

To prevent any damages of the device, ensure that no voltages are connected to pin 12! 

3.3  Connection to the PC 

If connecting 

meM-LOG-SE

 to a PC, it can be controlled via programming and measuring data stored in the logger 

memory can be read out. The connection is possible either via USB interface or via the serial interface RS232. 

Interface 

Description 

USB: 

-

 

fast transmission of measuring data 

-

 

to be used with operating systems Windows

®

 2000/XP/Vista 

serial (RS232): 

-

 

universal use on all operating systems supporting the serial interface, e. g. all Windows

®

 operating sys-

tems, Linux and Mac OS 

-

 

transmission over great distances, connection of a modem 

 

3.3.1 

Serial connection (RS232) 

The serial connection requires a 9-pole, cross-wired null modem cable, not electrically isolated, which is connected to 
the two 9-pole Sub-D plugs of the logger and the PC (to be ordered under: 

ZUKA-SER9

). The serial connection of the 

logger is located at the back of the device. If your PC only provides a 25-pole serial connection, you need an adapter 9-
pole to 25-pole in addition. The pin assignment of the 9-pole Sub-D plug for the serial interface is listed in the table 
below: 

Pin (Sub-D9) 

meM-LOG-SE 

2 RX 

3 TX 

5 GND 

(Ground) 

 

 

 

Programming of the meM-LOG-SE can only be done via the serial interface! 

 

Serial commands and data can only be sent between logger and PC if the transmission rates (baudrate) of 
the RS232 interface and the meM-LOG-SE are the same. 

3.3.2 USB 

If using the USB interface, connect the included USB cable 

ZUKA-USB

 to the USB connections both at PC and meas-

uring system. The socket for the USB interface is located at the front of the logger. 

3.4  Power supply connection 

The device is supplied with power via the 3-pole DIN plug at the back of the 
device. The input voltage must be in the range of 9..45V DC.  

External power supply is also required if using the USB interface of the 
PC!

 Optionally available as accessory is the power supply 

ZU-PW10W

 (12V, 

12W). 

5
4
3
2
1

9
8
7
6

GND

TX

RX

Summary of Contents for MEM-LOG-SE

Page 1: ... it is no problem to realize measuring applications which require mobility and flexibility in a high degree In order to extend the operating time of the accumu lator the device changes into an energy saving mode after 1 minute Sleep Mode and just wakes up occasionally e g shortly before scanning the next measuring value Via 16 analog input channels in the measuring range of 5V with a resolution of...

Page 2: ...ition and processing measuring files created with ST meM LOG SE can be dis played in NextView 4 as graphic signal curves and then be ana lyzed NextView 4 is also available as a stripped down Analysis version A free demo of the software is included with delivery The instruction set for program ming the meM LOG SE is com patible to the ADAM4000 series For further information please visit our homepag...

Page 3: ... each channel separately storage of either all channels to be saved or only of channels which are in state of alarm If measuring values are within their normal range again the scan is temporarily interrupted If not in Loop Mode see below the scan continues until memory is full or measurement is stopped even if a scan has been stopped with the device button it can be taken up again the new measurin...

Page 4: ...listed in the following table Pin Sub D15 meM LOG SE 1 Alarm Out H active high 9 Alarm Out L active low 8 15 Relay contact alarm 3 Digital In 1 active low 11 Digital In 2 active low 12 Scan start active low 5 13 Ground 2 4 6 7 10 14 n c In case of alarm the alarm outputs and the relay contact at pin 1 9 or 8 15 become active at the same time as long as the signal values exceed the defined range Al...

Page 5: ...ion over great distances connection of a modem 3 3 1 Serial connection RS232 The serial connection requires a 9 pole cross wired null modem cable not electrically isolated which is connected to the two 9 pole Sub D plugs of the logger and the PC to be ordered under ZUKA SER9 The serial connection of the logger is located at the back of the device If your PC only provides a 25 pole serial connectio...

Page 6: ...scan Æ turn off device measuring values are saved up until the point of turning off at least 5sec Scan running repeated and short flashing of LED fast flickering after 0 5sec after 5sec LED off Instant On temporarily inter rupt energy saving mode 0 5sec Sleep Mode not changing If you start a scan with the device button or a digital signal at the digital connector the memory must be empty This does...

Page 7: ... was put on user friendliness and reliability Administration of USB and the serial interface in one program The connection between PC and data logger is realized via the serial RS232 as well as the USB interface so that the advantages of both interfaces are available to the user see chapter 3 3 ST meM LOG SE manages both inter faces so that one device or several can be connected to USB and another...

Page 8: ... possible in all operating modes Store measuring data with different file format When reading out measuring data from the memory of the data logger ST meM LOG SE writes files in txt or lfx file format which can be displayed as table values and analyzed in various programs for word processing Notepad Wordpad Word etc and spreadsheet e g Excel Especially to mention is the storage as an lfx file the ...

Page 9: ... the software automatically SP2 do not connect with Windows Update Windows 2000 Search for a suitable driver for my device Under Windows Vista no selections have to be made The driver is found and installed automatically upon connection of the hardware because the driver package has been installed on hard disc before For the same reason no additional location needs to be entered for the driver sea...

Page 10: ...0Hz channel Sampling period 18h 12min 15sec 1sec in steps of sec 10min 55sec 350msec 0 01sec in steps of 1 100 sec Total sampling rate 16Hz 1 6kHz The total sampling rate is the sum of the sampling rates of the individual used channels e g 5 channels à 100Hz 500Hz total sam pling rate The values for accuracy always relate to the respective measuring range Errors might add at worst Digital inputs a...

Reviews: