2.5.3.2
BACKREST POSITION RESET
Perform only if the backrest emergency manual unlock (red handle) has been used.
Press and hold the “backrest” key and raise the backrest fully.
2.5.3.3
RESET FUN CTION
To restart the product following switching to safety mode after a power outage, over-current or
surge, press and hold the
“RAISE LEGREST” and “LOWER LEGREST”
keys until the audible
signal (repeated beeps) stops.
If the procedure fails, check the following connections:
-
Power cord
-
Battery cable
-
Fastening the 5 cables to the unit under the chair
-
Actuator cable under hard surface (less accessible)
-
Ensure that the module is powered on when connected to the mains.
Repeat the reset procedure from the start.
If the chair moves to an unwanted or extreme position, stop immediately to avoid damaging
the equipment. Contact the BMB MEDICAL Customer service for a repair intervention