BMB MEDICAL – CAREXIA – 1VCRX092020
78
EMISSIONS
Conducted and radiated RF emissions:
CISPR11 Group 1 Class A
Harmonic distortions IEC 61000-3-2:
Class A
Voltage fluctuations and Flicker IEC 61000-3-3:
Compliant
Immunity test
Test level IEC 60601
Compliance level
Electrostatic discharges (ESD)
IEC 61000-4-2
± 8 kV in contact
± 2/4/8/15 kV in air
± 8 kV in contact
± 2/4/8/15 kV in air
RF radiated electromagnetic fields
IEC 61000-4-3
3 V/m (professional healthcare establishment)
80 MHz-2.7 GHz
80% AM at 1 kHz
10 V/m
80MHz-2.7GHz
80% AM at 1 kHz
Proximity fields emitted by
wireless RF communication devices
IEC 61000-4-3
Frequency (MHz)
Modulation
Required level
(V/m)
Compliance level
(V/m)
385
Pulsed modulation: 18 Hz
27
27
450
Pulsed modulation: 18 Hz
28
28
710-745-780
Pulsed modulation: 217 Hz
9
9
810-871-930
Pulsed modulation: 18 Hz
28
28
1720-1845-1970
Pulsed modulation: 217 Hz
28
28
2450
Pulsed modulation: 217 Hz
28
28
5240-5500-5785
Pulsed modulation: 217 Hz
9
9
Rapid / burst electrical spikes
IEC 61000-4-4
Power supply: ± 2 kV
Input/output lines: ± 1 kV
Repeat frequency: 100 kHz
Power supply: ± 2 kV
Input/output lines: ± 1 kV
Repeat frequency: 100 kHz
Shock waves
IEC 61000-4-5
Between phases:
± 0.5 kV, ± 1 kV
Between earth and phases
± 0.5 kV, ± 1 kV, ± 2 kV
Between phases:
± 0.5 kV, ± 1 kV
Between earth and phases
± 0.5 kV, ± 1 kV, ± 2 kV
Conducted interference, induced by
RF fields
IEC 61000-4-6
3 V 0.15 MHz – 80 MHz
6 V in ISM bands between 0.15 MHz and 80
MHz
80 % AM at 1 kHz
3 V 0.15 MHz – 80 MHz
6 V in ISM bands between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
Power frequency magnetic fields
IEC 61000-4-8
30A/m
30A/m
Voltage Dips and Interruptions:
IEC 61000-4-11
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0% UT; 1 cycle at 0°
70% UT; 25/30 cycles at 0°
0% UT; 250/300 cycles
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% UT; 1 cycle at 0°
70% UT; 25/30 cycles at 0°
0% UT; 250/300 cycles
NB: There is no deviation from the requirements of IEC 60601-1-2 ed. 4
NB: There are no other known measures for maintaining basic safety with respect to EMC
phenomena.
Summary of Contents for CAREXIA FP
Page 1: ...BMB MEDICAL CAREXIA 1VCRX092020 1 CAREXIA MANUEL D UTILISATION USER MANUAL FRAN AIS ENGLISH...
Page 7: ...BMB MEDICAL CAREXIA 1VCRX092020 7 VUE D ENSEMBLE...
Page 9: ...BMB MEDICAL CAREXIA 1VCRX092020 9 VUE D ENSEMBLE...
Page 11: ...BMB MEDICAL CAREXIA 1VCRX092020 11 VUE D ENSEMBLE...
Page 13: ...BMB MEDICAL CAREXIA 1VCRX092020 13 VUE D ENSEMBLE...