AlcoTrue® E - Operating instructions | bluepoint® MEDICAL GmbH & Co. KG 2014
15
•
Measurement accuracy
Reproducibility with an ethanol standard
0 to 0.50 mg/L BrAC
≤
0,024 mg/L
0 to 1,00 ‰ BAC
≤
0,05 ‰
•
Ready for operation
After switching on
ca. 6 s
After completing a measurement in range (BrAC):
Typ.
0,00 mg/L
≤
2s
0,50 mg/L
≤
15s
After completing a measurement in range (BAC):
Typ.
0,00‰
≤
2s
1,00‰
≤
15s
•
Mouthpiece type
Typ „S“ from bluepoint Medical GmbH & Co. KG
•
Calibration interval
1000 measurements or 12 months
•
Calibration type
Wet gas
Permissible calibration range for unit of measuring value:
mg/L
0.24 to 0.71 mg/L
‰
0.50 to 1.50 ‰
mg/100mL
50 to 150 mg/100mL
Standard calibration value:
mg/L
0.435 mg/L
‰
1.00 ‰
mg/100mL
100 mg/100mL
The use of these standard calibration values corresponds to a conversion factor
of 2300:1 between blood and breath alcohol concentration values.