Blue-White industries flexflo A-100NV Operating Manual Download Page 10

A-100N

Page 10

5.2

OPERATING MODE 1 - Output adjusted manually - 

In this mode, the pump’s motor speed is adjusted manually using the front panel 
touch pad. The motor speed can be adjusted from 0-100%. To adjust the speed:

4

Set the pump for mode 1

. Press the MODE button until 

MODE 1

 is shown on 

the LCD display. The 

%SPEED

 icon will light. The large 

3-DIGIT LCD

 will 

indicate the currently programmed percentage of speed. 

4

Enter the programming mode.

 At the same time, press the RUN/STANDBY 

button and the MODE button. A blinking 

ARROW 

will point to the word 

PROGRAM 

indicating the program mode has been activated.

4

Press the FIELD button to select the digit to program. The digit will blink 
when selected.

4

Press the DIGIT button to change the selected digit.

4

Repeat until all digits are programmed.

4

To exit the programming mode, press the RUN/STANDBY button and the 
MODE button at the same time. The arrow next to the word 

PROGRAM 

will 

disappear.

[

NOTE: 

If while in the program mode no buttons are pressed within 20 seconds, 

the circuitry will automatically return to the run mode, without saving changes.

OPERATING MODE 2 - Output adjusted by 4-20 mA input signal 

In this mode, the pump’s motor speed is adjusted automatically based on the value 
of the 4-20 mA input signal. Any motor speed can be assigned to either the 
minimum or maximum milliamp input values. However, 

the programmed 

minimum mA value must be less than the programmed maximum mA value. 

The 

ALARM 

and 

SERVICE 

icons will blink if the programming is in error. To 

assign the minimum and maximum motor speed and the minimum and maximum 
mA input signal values:

4

Set the pump for mode 2

. Press the MODE button until 

MODE 2

 is shown on 

the LCD display. The 

%SPEED 

or 

mA 

icon will light depending on the 

current display setting. The large 

3-DIGIT LCD

 will indicate the current 

motor speed or the current mA input value. 

4

Enter the programming mode.

 At the same time, press the RUN/STANDBY 

and MODE buttons. A blinking 

ARROW 

will point to the word 

PROGRAM 

indicating the program mode is activated. A blinking 

ARROW 

will point to the 

word 

MINIMUM 

indicating the minimum value is ready to be programmed. 

The 

% SPEED

 icon will blink indicating the percentage of speed is ready to 

be programmed.

4

Enter the motor speed at the minimum mA input signal value.

 Press the 

FIELD button to select the digit to program. The digit will blink when 
selected.

4

Press the DIGIT button to change the selected digit.

4

Repeat until all digits are programmed.

4

Press the mode button. The 

% SPEED

 icon will stop blinking and the 

mA 

icon 

will blink indicating the minimum mA value is ready to be programmed. The 
currently programmed minimum value is shown on the 

3-DIGIT LCD

.

4

Enter the minimum mA input signal value.

 Note: this value must be less than 

the maximum mA input signal value. Press the FIELD button to select the 
digit to program. The digit will blink when selected.

4

Press the DIGIT button to change the selected digit.

4

Repeat until all digits are programmed.

5.3

 - 

Slide both top & bottom clamps to the left only far enough

(as shown) to open the control cover

5.0

How To Operate The A-100NV

5.1

Description of Pump Output Adjustment Controls  - 

Open the control panel door by sliding the upper and lower slide 
clamps to the left.  FIG. 5.2

!

RUN/STANDBY 

Button - 

4

Press to start and stop the pump. The 

ARROW 

next to the word 

RUN 

will light when in the run mode. The 

ARROW 

next to the word 

STAND-BY 

will blink when in the stand-by mode.

4

Press to clear 

ALARM.

4

When pressed with the FIELD Button,  initiates a 99 second prime 
cycle which temporarily overrides the mode setting and runs the 
pump motor at 100% speed. The 

ARROW 

next to the word 

PRIME 

will blink.

4

When pressed with the DIGIT button, resets the 500 hour service 
warning timer to zero.

4

When pressed with the MODE button, initiates the programming 
mode. The 

ARROW 

next to the word 

PROGRAM 

will blink.

!

FIELD 

Button - 

4

In the programming mode, selects the digit to be changed.

!

DIGIT 

Button - 

4

In the programming mode, increases the selected digit. 

4

When pressed with the MODE Button,  toggles the display from % 
motor speed to input signal value.

!

MODE 

Button - 

4

Used to select one of four operating modes.

Mode 1 

- Manual Adjustment (external input disabled)

Mode 2 

- 4-20mA input

Mode 3 

- 0-10VDC input

Mode 4 

- Frequency (Hz) input 

FIG. 5.2

FIG. 5.1

  SLIDE CLAMP

RUN

STANDBY

RUN

FIELD

DIGIT

MODE

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

INPUT MODES

1 - MANUAL

2 - 4-20mA

3 - 0-10VDC

4 - PULSE (Hz)

PROGRAM

RESET SERVICE

PRIME

DISPLAY

VARIABLE SPEED PUMP

% SPEED

1

MODE

VDC

SERVICE

1000

ALARM

mA

Hz

RUN

STANDBY

RUN

FIELD

DIGIT

MODE

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

INPUT MODES

1 - MANUAL

2 - 4-20mA

3 - 0-10VDC

4 - PULSE (Hz)

PROGRAM

RESET SERVICE

PRIME

DISPLAY

VARIABLE SPEED PUMP

% SPEED

1

MODE

VDC

SERVICE

1000

ALARM

mA

Hz

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

% SPEED

1

MODE

100

RUN MODE 1

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

% SPEED

1

MODE

000

PROGRAM MODE 1

constant speed % setting

RUN MODE 2

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

2

MODE

04.0

mA

PROGRAM MODE 2

% speed at the minimum input

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

% SPEED

2

MODE

000

PROGRAM MODE 2

minimum input value

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

2

MODE

04.0

mA

A-100N

Page 11

Summary of Contents for flexflo A-100NV

Page 1: ...le Speed Peristaltic Injector Pump Operating Manual 5300 Business Drive Huntington Beach CA 92649 USA Phone 714 893 8529 FAX 714 894 9492 E mail sales blue white com or techsupport blue white com Webs...

Page 2: ...o maintain the A 100N 16 7 1 Routine inspection and cleaning 16 7 2 How to clean and lubricate the A 100N 16 7 3 500 hour service warning timer 16 7 4 How to replace the pump tube 17 Replacement parts...

Page 3: ...system 2 0 Specifications Maximum Working Pressure 100 psig 6 9 bar o o Maximum Fluid Temperature 130 F 54 C o o Ambient Temperature Range 14 to 110 F 10 to 43 C Duty Cycle Continuous Maximum Solids 5...

Page 4: ...the pump tube Do not mount the pump directly over your chemical container Chemical fumes may damage the unit Mount the pump off to the side or at a lower level than the chemical container Mounting th...

Page 5: ...umping unit R Suction Tube Chemical Container NPT Injector Page 5 RUN STANDBY RUN FIELD DIGIT MODE PROGRAM STAND BY PRIME MINIMUM MAXIMUM INPUT MODES 1 MANUAL 2 4 20mA 3 0 10VDC 4 PULSE Hz PROGRAM RES...

Page 6: ...ections The A 100NV will accept any one of three different types of external input signals 4 20 mA 0 10 VDC or frequency Two types of frequency inputs AC sine waves magnetic coils type outputs and Dig...

Page 7: ...WHITE frequency input JUNCTION BOX LIQUID TIGHT CONNECTOR ACCEPTABLE CABLE JACKET RANGE 118 255 INCH 3 0 6 5 MM EXTERNAL INPUT CABLE BROWN motor on contact output INPUT SIGNAL CONNECTOR 7 wire bundle...

Page 8: ...e tube Secure the ceramic weight to the strainer Drop the strainer into the solution tank 4 3 3 Outlet Tubing Locate the outlet fitting of the Pump Tube see fig 4 6 Remove the tube nut Push the opaque...

Page 9: ...he back pressure and interfering with the check valve operation See section 7 0 Install the Injection Check valve directly into the piping system Do not use a pipe stud with a tee for insertion of the...

Page 10: ...pressed with the MODE button initiates the programming mode The ARROW next to the word PROGRAM will blink FIELD Button 4 In the programming mode selects the digit to be changed DIGIT Button 4 In the...

Page 11: ...mode 2 Press the MODE button until MODE 2 is shown on the LCD display The SPEED or mA icon will light depending on the current display setting The large 3 DIGIT LCD will indicate the current motor spe...

Page 12: ...LD button to select the digit to program The digit will blink when selected 4 Press the DIGIT button to change the selected digit 4 Repeat until all digits are programmed 4 Press the mode button Progr...

Page 13: ...maximum VDC input signal value Press the FIELD button to select the digit to program The digit will blink when selected 4 Press the DIGIT button to change the selected digit 4 Repeat until all digits...

Page 14: ...ut signal value Press the FIELD button to select the digit to program The digit will blink when selected 4 Press the DIGIT button to change the selected digit 4 Repeat until all digits are programmed...

Page 15: ...in the bottom of the pump head just enough to cover the sensors and turn on the pump If the TFD system detects the chemical the pump will stop after a five second confirmation period and the ALARM ic...

Page 16: ...may require grease periodically Apply a small amount of grease Aeroshell aviation grease 5 or equivalent when necessary Silicone oil ONLY may be used on the roller assembly and tube assembly The inje...

Page 17: ...on side of the Pump Tube fitting into the lower retaining slot in the pump head Fig 7 2 6 4 2 2 Carefully guide the Pump Tube into the pump head Stretch the tube slightly and insert the outlet dischar...

Page 18: ...0 15 16 11 3 14 13 31 30 17 26 27 3 2 1 2 9 52 54 5 4 7 8 56 55 10 12 28 29 25 24 23 22 21 33 32 34 F IE L D 6 53 D IG IT M O D E ST AN DB Y PR IM E R ES ET 99 SE C C YC LE TU BE LIF E WA RIN G TIM ER...

Page 19: ...8 Star 1 25 90010 222 Wire Motor ground Digital Timers Green 1 26 A 008 1 Gearbox 14 Rpm 1 A 008 2 Gearbox 30 Rpm 1 A 008 3 Gearbox 45 Rpm 1 A 008 4 Gearbox 60 Rpm 1 A 008 5 Gearbox 125 Rpm 1 27 76001...

Page 20: ...a brief description of the problem as well as the original invoice or sales receipt showing the date of purchase The receipt will be returned with the unit Prepay all shipping costs COD shipments will...

Reviews: