Key Points and Precautions for Use
Maintenance must be carried out by competent personnel.
It is recommended that maintenance is carried out annually and a service carried out as a
condition of warranty every 20 months.
A regular visual check of the foot is recommended, signs of wear or corrosion that may affect
function should be reported to the service provider.
Key Points
Any changes in performance of this device must be reported to the practitioner
Changes in performance may include:
• Increase in ankle stiffness
• Reduced ankle support (freer movement)
• Any unusual noise (silent buzz from the device during charging is normal)
• Recurring rapid flashing of Charging Indicator/Status Indicator LEDs.
At low battery power levels the device will go into sleep mode, switch off dynamic control and
use only the basic resistance settings and may affect
Bluetooth®
connectability. Charge the
device batteries as soon as possible to restore Dynamic mode.
The practitioner must also be informed of the following:
• Any changes in body weight and/or activity level.
• Significant wear or excessive discoloration from long term exposure to UV.
Cleaning
Use a damp cloth and mild soap to clean outside surfaces, DO NOT use aggressive
cleansers.
Transport and Storage
Storage temperature range between -15˚C and 50˚C (5 ˚F and 122 ˚F).
If the device is shipped, send the device and the charging pod in the packaging supplied with
the unit. Always ship the device with the charger pod.
When built into a limb it should be stored securely so as not to damage the unit.
Lifting Loads:
User weight and activity is governed by the stated limits.
Load carrying by the user should be based on a local risk assessment.
Environment
This device is IP67 rated: waterproof to a maximum
depth of 1 meter. Occasional submersion should be
limited to a max. of 30 minutes. Dry after using in
water.
Avoid exposing the device to abrasive or corrosive environments, such as those containing sand,
acids, salt, chlorinated, or chemically treated water e.g. swimming pools. Failure to comply will
void the warranty. After any accidental exposure to such environments rinse the device in fresh
water and dry thoroughly.
Exclusively for use between -15˚C and 50˚C (5˚F and 122˚F).
Charge only between 0 ˚C to 35 ˚C (32 ˚F to 95 ˚F)