Programming Using Pulse Output Turbine
Flow Meters
Each turbine flow meter is shipped with either a K-factor value or
frequency data. If frequency data is provided, the data must be
converted to a K-factor before programming the monitor. K-
factor information, when supplied, can usually be found on the
neck of the flow meter or stamped on the body. The K-factor
represents the number of pulses per unit of volume. The K-factor
will be needed to program the monitor readout.
ENTER PROGRAMMING MODE
– Change to program-
ming mode by pressing the MENU button once. The mode
indicator will change from RUN to PROGRAM.
SELECT THE METER SIZE
– At the METER prompt,
press the UP or RIGHT arrow keys to select the bore size of
your meter. Press ENTER button once to save meter size
choices and advance to the K-factor units selection.