ADDITIONAL SCALING PAOAMETEOS

OTEE:

N

The programming instructions below are only available for loop powered units  Battery powered units do not include 

these programming parameters 

Flow 4 mA Setting

Zero is the default flow rate at the 4 mA setting  If the current selection is correct, press 

ENTEO

 to advance to the 

Flow 20 mA

 

parameter  To change the setting, press 

UP

 to increment the flashing digit and press 

OIGHT

 to move to the next digit  Press 

ENTEO

 to save and advance to the 

Flow 20 mA

 parameter 

Flow 20 mA Setting

The turbine meter’s maximum flow rate is the default value for the 20 mA setting  If the current selection is correct, press 

ENTEO

 to advance to the 

4…20 mA Calibration

 parameter  To change the setting, press 

UP

 to increment the flashing digit and 

press 

OIGHT

 to move to the next digit  Press 

ENTEO

 to save and advance to the 

4…20 mA Calibration

 parameter  

4…20 mA Calibration

The 

4…20 mA Calibration

 parameter is the fine adjustment of the 4…20 mA output  The 4 mA setting is typically between 

35…50  To set the 4 mA value, connect an ammeter in series with the loop power supply  At the 

4MA OUT

 prompt, press 

UP

 to 

increment and 

OIGHT

 to decrement the 4 mA value to until a steady 4 mA reading is reached on the ammeter  Press 

ENTEO

 to 

lock in this value and advance to 

20MA OUT

  The 20 mA adjustment is performed using the same procedure as the  

4 mA adjustment 

4…20 mA Test

The monitor contains a diagnostic routine that allows the simulation of mA values between 4…20 to check output tracking  

At the 

4-20TEST

 prompt press 

UP

 to increment the simulated mA output or 

OIGHT

 to decrement the value in increments of  

1 mA  The ammeter should track the simulated mA output  If a 4…20 mA test in not necessary, press 

ENTEO

 once to escape 

the testing at any time 

Pulse Out

Freq. In

4-20 mA

10…30V DC

 

1…5V DC 

1

2

3

4

5

6

+          -

+         -

+          -

JP2

Mag

Input

Pulse

Inuput

TB1

1 2 3 4 5

J2

Keypad

Connector

DC

PULSE OUTPUT

IO/SETUP

DSP

Flow 4MA

Flow 20MA

4MA  OUT

20MA  OUT

4-20 TEST

PASSWORD

Figure 4:  Typical ammeter connection

Figure 5:  4…20 mA programming menu

Additional Scaling Parameters

Page 8 

March 2017 

DSY-PM-00278-EN-03

Summary of Contents for B2800 SERIES

Page 1: ...Flow Monitor B2800 Panel Mount Standard DSY PM 00278 EN 03 March 2017 User Manual...

Page 2: ...5 Modes 5 Programming 6 Programming Mode 6 Run Mode 7 Additional Scaling Parameters 8 Flow 4 mA Setting 8 Flow 20 mA Setting 8 4 20 mA Calibration 8 4 20 mA Test 8 Battery Replacement 9 Additional Inp...

Page 3: ...serious personal injury Indicates a hazardous situation which if not avoided could result in severe personal injury or death Indicates a hazardous situation which if not avoided is estimated to be ca...

Page 4: ...umeric display provides for selectable units viewing in run mode and prompts for variables in program mode The display can show rate total or alternate between both rate and total Figure 1 B2800 Panel...

Page 5: ...0 panel mount flow monitor Buttons MENU Switches to PROGRAM mode UP Scrolls forward through the parameter options and increments numeric variables RIGHT Scrolls backward through the parameter options...

Page 6: ...and the bore size are different For example many of the one inch NPT turbines have bore sizes that range from 3 8 1 in Be sure to use the correct bore size or the meter will report incorrect flows and...

Page 7: ...e Initially the password is set to all zeros and any user can modify the parameter settings in PROGRAM mode To change the password enter any four digit code Press UP to increment the digit and RIGHT t...

Page 8: ...the 4 20 mA output The 4 mA setting is typically between 35 50 To set the 4 mA value connect an ammeter in series with the loop power supply At the 4MA OUT prompt press UP to increment and RIGHT to de...

Page 9: ...ct closure input pulse Since most Blancett turbine flow meters utilize a magnetic pickup the B2800 flow monitor is shipped configured for magnetic pickup input To change to a contact closure input rem...

Page 10: ...perature Drift 50 ppm C maximum Mounting Style Panel Mount 3 6 inch 3 6 inch Opening NEMA 4 front only Environmental Operating Temperature 22 158 F 30 70 C Humidity 0 90 Non condensing Inputs Type Mag...

Page 11: ...s reading constantly This is usually an indication of external noise Keep all AC wires separate from DC wires Check for large motors close to the meter pickup Check for radio antenna in close proximit...

Page 12: ...tandard The Americas Badger Meter 4545 West Brown Deer Rd PO Box 245036 Milwaukee WI 53224 9536 800 876 3837 414 355 0400 M xico Badger Meter de las Americas S A de C V Pedro Luis Ogaz n N 32 Esq Ange...

Reviews: