11
Operations & Maintenance Manual: Introduction
SCREEN ICONS
The following icons may appear on the instrument LCD screen.
Table 4.
Probe screen icons
ICON
MEANING
Battery power level.
Female gender option is selected. Select this option only for women who have not
had a hysterectomy. Deselect for all others, male or female.
Bladder imaging in progress. Hold instrument steady.
Solid
: Indicates that the bladder was not centered within the ultrasound field of view.
However, the bladder volume measurement is still accurate. Re‑aiming is optional.
Flashing
: Indicates that the aim was “off target.” In order to get an accurate bladder
volume measurement, you must re‑aim the probe in the direction of the arrow.
The patient’s actual bladder size is larger than the ultrasound field of view.
Indicates the number of days remaining until the next required calibration.