Operations & Maintenance Manual: Important Information
The charging cradle, power adapter, and power cords are not intended for patient contact.
Ensure 2 m (6 ft) is maintained between the patient and these components.
WARNING
Ensure proper distance from patient. When transmitting data to or from your computer, make sure
the instrument, accessories, and computer are outside the patient vicinity (more than 2 m (6 ft)
from the patient).
WARNING
Do not use the system on:
• Fetal patients.
• Pregnant patients.
• Patients with open skin or wounds in the suprapubic region.
• Patients with ascites.
WARNING
To reduce the risk of electric shock or burns, do not use the system in conjunction with
high‑frequency surgical equipment.
WARNING
To reduce the risk of electrical shock, do not attempt to open the system components. This may
cause serious injury to the operator or damage to the instrument and will void the warranty.
Contact Verathon
®
Customer Care or your local representative for all servicing needs.
WARNING
To reduce the risk of explosion, do not use the system in the presence of flammable anesthetics.
WARNING
No modification of this equipment is allowed.