Warnings and Precautions
Warnings
Do not touch any exposed electrical conductors when the CerePlex µ
connector is attached to devices on subject’s head as this may result in
inducing electric charge to the neural tissue. Irreversible damage may
occur.
Do not use the CerePlex µ in the presence of flammable anesthetic
agents or any other reagents.
Avoid strong static discharges from sources like television or computer
monitors because they can damage the electrical components of the
system.
Keep the CerePlex µ away from liquids. Contact with water, shower
spray, or wet surfaces can lead to the subject receiving an electrical
shock.
Connection of external instruments to the CerePlex µ may compromise
electrical safety.
Always use antistatic or electrostatic discharge (ESD) safe gloves when
connecting the CerePlex µ.
Use only the supplied Blackrock Microsystems components (Cerebus™
system & Digital Hub128 or CerePlex Direct, CerePlex µ Cable).
Substitution of components not supplied by Blackrock Microsystems may
affect system performance and subject safety.
Use caution when placing cables and other connectors to minimize the
likelihood of tripping or accidentally pulling on cables. Pulled cables may
cause damage to the CerePlex µ or any other connected devices.
Precautions
Read this entire manual prior to using the device.