Pull a transducer cover or a standard condom containing a suitable amount of
imaging gel over the transducer.
3
Slide the needle guide into the biopsy channel bracket.
4
Insert the assembled needle guide and biopsy channel bracket into the open
channel on the transducer. A small nodule on the end of the channel bracket fits
into an indentation in the channel on the transducer to help you place the bracket
correctly. Click the channel bracket into position on the transducer and lock it
into place (see Figure 11).
Figure 11. Biopsy channel bracket and needle guide mounted on 8818.
For Transperineal Puncture
The metal puncture attachment UA1324, shown in Figure 12, is designed for
transperineal puncture and biopsy. When UA1324 is being used, the dummy channel
bracket UA1325 (shown in Figure 3) must be in place.
WARNING
Do not use excessive force when inserting the needle guide.
WARNING
Ensure that the channel bracket and needle guide are correctly positioned. Never insert the
needle guide while the transducer is inside the patient.