12
March
2017
8818 User Guide
(BB1505-H)
Please refer to
Care and Cleaning
for an example of how to open a sterile-packed
product.
Simultaneous Biplane
The sterile single-use biopsy guide UA1322-S, UA1322-S14 and reusable biopsy
guide UA1326 are used for simultaneous biplane imaging.
Figure 5. Reusable biopsy guide UA1326.
The puncture line for UA1322-S, UA1322-S14 and UA1326 on 8818 is shown in
Figure 6. UA1322-S and UA1326 have a bore diameter of 1.6mm, suitable for 18-
gauge needles. UA1322-S14 has a bore diameter of 2.1mm, suitable for 14-gauge
needles.The puncture line is angled at 19° to the transducer’s axis.
Figure 6. Illustration of the puncture line for biopsy guide UA1322-S, UA1322-S14 and UA1326.
Endfire
The sterile single-use biopsy guide UA1323-S and reusable biopsy guide UA1327
are used in endfire imaging.
Figure 7. Reusable biopsy guide UA1327.
WARNING
For contaminated disposables such transducer covers or needle guides, follow disposal
control policies established for your office, department or hospital.