Byte
offset
Meaning
3
Calibration protection state:
Bit 0: value of 0 means the calibration state is not protected
Bit 0: value of 1 means the calibration state is protected
4-24
Reserved
0x62 Set voltage calibration point index
Byte
offset
Meaning
3
Voltage calibration point. Valid values are 0x01 to 0x04.
4-24
Reserved
The DC Load has four calibration points for voltage. These would typically span the full range of the
instrument's voltage range.
0x63 Send the actual voltage to the calibration program
Byte
offset
Meaning
3 to 6
4 byte little-endian integer representing the voltage in units of mV
7-24
Reserved
0x64 Set current calibration point index
Byte
offset
Meaning
3
Current calibration point. Valid values are 0x01 to 0x04.
4-24
Reserved
The DC Load has four calibration points for current. These would typically span the full range of the
instrument's current range.
0x65 Send the actual current to the calibration program
Byte
offset
Meaning
3 to 6
4 byte little-endian integer representing the current in units of 0.1 mA
7-24
Reserved
0x66 Store the calibration data to EEPROM
Byte
offset
Meaning
3-24
Reserved
8500 DC Load Series
Version: February 4, 2009
Page 68 of 76