OPERATING INSTRUCTIONS

 

  636

INFRARED THERMOMETER

WITH LASER SIGHTING

INTRODUCTION

This instrument is a portable easy to use 3½digit,
compact-sized digital infrared thermometer with laser
sighting designed for simple one hand operation. Meter
comes with Backlit LCD display,  Auto-Hold function and
auto power down (10 seconds approx.) after releasing
Trigger to extend battery life.

SAFETY  INFORMATION

It is recommended that you read the safety and operation
instructions before using the infrared thermometer.

CAUTION

• Do not use the unit near any device which generates

strong electromagnetic radiation or near a static

electrical charge, as these may cause errors.

• Do not use the unit where it may be exposed to

corrosive or explosive gases. The unit may be dam-

aged, or explosion may occur.

• Do not keep or use this unit in an environment where

it will be directly illuminated by sunshine, or where it

will be exposed to high temperatures, high humidity or

condensation. If you do, it may be deformed, its

insulation may be damaged, or it may no longer

function according to specification.

• Do not point the lens at the sun or at any other source

of strong light. If you  do, the sensor may be damaged.

• Do not contact the lens against the object whose

temperature is to be measured, or get it dirty, allow it

to be scratched, or allow any foreign material to

adhere to it. Doing so may cause errors.

• Do not touch  or hold by the front case. Temperature

reading can be affected by heat from hand.

• Do not place the meter on or around hot objects (70°C/

158°F). It may cause damage to the case.

• If the meter is exposed to significant changes in

ambient temperature (hot to cold or cold to hot). Allow

20 minutes for temperature stabilization, before tak-

ing measurement.

• Condensation may form on the lens when going from

a cold to hot environment-wait 10 minutes for conden

sation to dissipate before taking measurements.

• This unit is not constructed to be water proof or

dustproof, so do not use it in a very dusty environment

or in one where it will get wet.

SPECIFICATIONS

GENERAL

Display:

3½ digit liquid crystal display (LCD) with maximum
reading of 1999

Low battery indication: 

 the " 

 " is displayed when

the battery voltage drops below the operating level

Measurement rate: 

0.25 second, nominal.

Operating Environment: 

32°F to 122°F (0°C to 50°C)

at < 70% R.H.

Storage Temperature:

-4°F to 140°F (-20°C to 60°C)  , 0 to 80% R.H. with
battery removed from meter

Auto power off:

 10 seconds.

Standby consuming current:

 <5µA

Battery:

 Standard 9V battery (NEDA 1604, IEC 6F22

006P)

Battery Life:

9 hours (continuity) typical
(contain Laser and Backlit)

Dimensions(HxWxD): 

5.8x4.1x1.65"(148 x 105 x

42mm).

Weight:

 approx. 157g  (including battery.)

Laser Specifications

Laser safety classification of Class 2
Wave Length:

 Red (630~670nm).

Power out:

 <1mW, class 2 laser product.

 DANGER

Pressing the Trigger turns the laser beam on and off.
Exercise extreme care and do not allow the laser beam
to enter your eye or those of any other person or animal.
• Do not look directly into the laser light from the optical

system.

• When measuring the temperature of an object which

has a mirror finish, be careful not to allow the laser
light beam to be reflected off the surface into your eyes
or those of another person.

• Do not allow the laser light beam to impinge upon any

gas which can explode.

EMC/RFI

Readings may be affected if the unit is operated within a
radio frequency electromagnetic field strength of approxi-
mately 9 volts per meter, but the performance of the
instrument will not be permanently affected.

®

ELECTRICAL

Temperature Range:

 -22°F to 1022°F / -30°C to 550°C

Display Resolution:

 1°F, 0.5/1°C (Auto)

Accuracy:

±(4°F/2°C)  for  -22°F to 212°F, -30°C to 100°C

±(2% reading) for  213°F to 1022°F, 101°C to 550°C

Temperature Coefficient:

 ±0.2% of reading or ±0.36°F/

0.2°C, whichever is greater, change in accuracy per °F/

°C change in ambient operating temperature above 82.4°F/

28°C or below 64.4°F/18°C.

Response Time: 

0.25 second

Spectral Response:

 6 to 14µm nominal

Adjustable emissivity (

εεεεε

):

 0.1 to 1.0

Detection Element:

 Thermopile

Optical Lens:

 Fresnel Lens

Sighting:

 1-beam laser marker <1mW (class 2)

Field of View:

 100mmØ at 1000mm (3.9"Ø at 39.0")

Spot size increases with distance from the probe tip as shown

(Spot Diameter measured at 90 % Energy)

OPERATING INSTRUCTIONS

Trigger

Pull the trigger to turn on the meter for measuring
temperature. Releasing trigger to stop measuring tem-
perature and automatically hold the display
reading, the meter power down automatically after 10
seconds.

Button function:

1. Set button

SET MODE & Numeric input key

"SET" annunciator appears when a numerical value can be
set ( during setting of 

ε

, ALM Hi and ALM Lo).

 key: The numerical value is increased.
 key: The numerical value is reduced.
If either of these numerical value keys is held down, the

numerical value changes rapidly in the appropriate
direction.

The data will store in nonvolatile storage region while

after  completing settings. In this mode the automatic power
-off feature is disable.

2. 

"   " button

Press " " button to on the " " annunciator. If " "

annunciator is on, press trigger and the laser beam will turn
on and " " annunciator will blink. Releasing trigger to turn
off the laser beam.

3. 

"    "  button

Use 

"    "

 button to select turn on or  off the Back-Light

function.

4. 

°C/°F button

Readings are displayed in either degrees Celsius(°C) or

degrees Fahrenheit(°F). When the thermometer is turned
on. To change the temperature scale by pressing °C/°F
button .

5. 

APO button

It will auto power off for about 10 seconds.

Press "APO" button to disable . Auto Power-Off function
that HOLD indication disappears and press again to enable
APO function.

Press SET button switches the mode around the cycle 

ε

 

!

ALM Hi 

!

 ALM Lo 

!

 MAX 

!

 MIN 

!

 HOLD.

HOLD

:Releasing the trigger to stop measurement of
temperature, the HOLD indication appears,
and the measured temperature is held.

ε

:The thermal emissivity of the object set using
the   and   keys. (refer to Table 1)

ALM Hi :The upper limit alarm temperature is set  using

the   and   keys. When the measured
temperature is exceeded the Hi setpoint, the

beeper emits a discontinuous pulse tone and
"ALM Hi" is displayed.

ALM Lo :The lower limit alarm temperature is set  using

the   and   keys. When the measured
temperature is below the Lo setpoint, the beeper
emits a continuous pulse tone and "ALM Lo"
is displayed.

MAX

:The maximum temperature during measure-

ment is displayed.

MIN

:The minimum temperature during measure-

ment is displayed.