background image

9

With the set up of 2.7.1, push the offset control. Rotate the offset knob from full CCW to full CW
and observe a waveform change from negative to positive level. The maximum DC plus peak is
limit to ±10V (un-terminated) or ±5V (50 ohms terminated).

2.7.3 Frequency

Connect a counter to the instrument output. Set the instrument controls as follows:

WAVEFORM ...................SINE
OFFSET ..........................OFF
ATTENUATOR................OUT
AMPLITUDE ...................MIDRANGE
DUTY CYCLE .................OFF

Change the frequency and range and read the counter indications. The reading should fall between
instrument specifications.

2.7.4 Sync Output

Connect the scope to the SYNC output. Use the set up of paragraph 2.7.1. Observe a square wave,
of no less than 2V amplitude, and with 50% symmetry.

2.7.5 Symmetry

With the set up of paragraph 2.7.1, select Square waveform and push the DUTY CYCLE ON
control. Rotate the DUTY CYCLE knob from full CCW to CW and observe a waveform symmetry
change .

Section 3

Operating Instructions

3.1 CONTROLS AND CONNECTORS

Front panel controls and connectors are shown in the front view illustration of the Model 4013.
Each operating control is described as follows:

1.

POWER ON/OFF – Applies or removes AC power to the unit.

2.

FREQUENCY – Selects the frequency of the output waveform.

3.

RANGE SWITCHES – Multiply or divide by 10 the selected frequency for faster

setting.

4.

WAVEFORM – Select Sine, Triangle and Square output.

5.

OFFSET ON/OFF – Enables operation of the offset control.

6.

OFFSET – Sets the DC offset level of the output waveform. The maximum peak plus DC
level is limited to ±10V (non terminated) or clipping of the waveform will occur.

7.

OUTPUT LEVEL – Varies the amplitude of the selected waveform. The range is over 20dB.

Summary of Contents for 4013DDS

Page 1: ...1 Model 4013DDS 12 MHz Sweep Function Generator...

Page 2: ...r consequential damages So the above limitation or exclusion may not apply to you This w arranty gives you specific rights and you may have other rights w hich vary from state to state Service Informa...

Page 3: ...1 2 DESCRIPTION 5 1 3 SPECIFICATIONS 5 1 4 TECHNICAL SUPPORT 6 Section 2 Installation 2 1 INTRODUCTION 7 2 3 INSTRUMENT MOUNTING 7 2 4 POWER REQUIREMENTS 7 2 5 GROUNDING REQUIREMENTS 7 2 7 ACCEPTANCE...

Page 4: ...presence offlammable gases or fumes Operation ofany electrical instrument in such an environment constitutes a definite safety hazard KEEP AWAY FROM LIVE CIRCUITS Instrument covers must not be remove...

Page 5: ...odel 4013DDS Function Generator 1 2 DESCRIPTION The Model 4013DDS is a versatile high performance function generator Implemented using advance digital technologies based on DDS topology the units gene...

Page 6: ...ariable up to 10V 5V into 50 Absolute peak amplitude plus of f set limited to 10V 5V into 50 OUTPUT IMPEDANCE 50 OUTPUT PROTECTION The generator main output is non destructiv ely protected against sho...

Page 7: ...eck for the Model 4013DDS Function Generator 2 2 MECHANICAL INSPECTION This instrument was carefully inspected before shipment Upon receipt inspect the instrument for damage that might have occurred i...

Page 8: ...cable must be plugged into an approved three conductor electrical outlet WARNING Do not alter the ground connection Without the protective ground connection all accessible conductive parts including c...

Page 9: ...set up of paragraph 2 7 1 select Square waveform and push the DUTY CYCLE ON control Rotate the DUTY CYCLE knob from full CCW to CW and observe a waveform symmetry change Section 3 Operating Instructi...

Page 10: ...specifications An RG 58 coaxial cable equipped with BNC connectors and terminated with a 50 feed through termination is recommended Using the 50 terminator minimizes signal reflection and power losses...

Page 11: ...ly press the 20 dB attenuator button and rotate the AMPLITUDE control to verify that the waveform amplitude changes Return these controls to the preset condition 5 Push the OFFSET ON button and rotate...

Page 12: ...not exceed 10V into open circuit and 5V into 50 The maximum current which can be taken from the output in DC or AC modes is limited to 100 mA maximum 3 5 5 Sweep Operation The internal sweep mode allo...

Page 13: ...13...

Reviews: