MAINTENANCE

WARNING

The following instructions are for use by
qualified service personnel only. To avoid
electrical shock, do not perform any serv-
icing other than contained in the operat-
ing instructions unless you are qualified
to do so.

High voltage up to 2000 V is present when
covers are removed and the unit is oper-
ating. Remember that high voltage may
be retained indefinitely on high voltage
capacitors. Also remember that ac line
voltage is present on line voltage input
circuits any time the instrument is
plugged into an ac outlet, even if turned
off. Unplug the oscilloscope and dis-
charge high voltage capacitors  before
performing service procedures.

FUSE REPLACEMENT

If the fuse blows, the “ON” indicator will not light and the

oscilloscope will not operate. The fuse should not normally
open unless a problem has developed in the unit. Try to
determine and correct the cause of the blown fuse, then
replace only with the correct value fuse. For 110/125 V line
voltage operation, use an 800 mA, 250 V fuse. For 220/240
V line voltage operation, use a 600 mA, 250 V fuse. The fuse
is located on the rear panel adjacent to the power cord
receptacle.

Remove the fuseholder assembly as follows:

1. Unplug the power cord from rear of scope.

2. Insert a small screwdriver in fuseholder slot (located

between fuseholder and receptacle). Pry fuseholder
away from receptacle.

3. When reinstalling fuseholder, be sure that the fuse is

installed so that the correct line voltage is selected (see
LINE VOLTAGE SELECTION).

LINE VOLTAGE SELECTION

To select the desired line voltage, simply insert the fuse

and fuse holder so that the appropriate voltage is pointed to
by the arrow. Be sure to use the proper value fuse (see label
on rear panel).

PERIODIC ADJUSTMENTS

Probe compensation and trace rotation adjustments

should be checked periodically and adjusted if required.
These procedures are given below.

Probe Compensation

1. Connect probes to

CH 1

and

CH 2

input jacks. Per-

form procedure for each probe, one probe at a time.

2. Set the probe to X10 (compensation adjustment is not

possible in the X1 position).

3. Touch tip of probe to

CAL

terminal.

4. Adjust oscilloscope controls to display 3 or 4 cycles of

CAL

square wave at 5 or 6 divisions amplitude.

5. Adjust compensation trimmer on probe for optimum

square wave (minimum overshoot, rounding off, and
tilt). Refer to Fig. 11.

Trace Rotation Adjustment

1. Set oscilloscope controls for a single trace display in

CH 1

mode, and with the channel 1

AC-GND-DC

switch set to

GND

.

2. Use the channel 1

POSition control to position the

trace over the center horizontal line on the graticule
scale. The trace should be exactly parallel with the
horizontal line.

3. Use the

TRACE ROTATION

adjustment on the front

panel to eliminate any trace tilt.

Correct

Compensation

Over

Compensation

Insufficient

Compensation

Fig. 11. Probe Compensation Adjustment.

20

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for 2120C

Page 1: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return path 5 When using a probe touch only the...

Page 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 4: ...ring 13 Magnified Sweep Operation 15 OPERATING INSTRUCTIONS Continued X Y Operation 15 Video Signal Observation 15 Application Guidebook 15 Delayed Sweep Operation 15 Component Test Operation 16 MAINT...

Page 5: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 6: ...Y Operation Channel 1 can be applied as horizontal deflection X axis while channel 2 provides vertical deflection Y axis Built in Probe Adjust Square Wave A 2 V p p 1 kHz square wave generator permit...

Page 7: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 8: ...equency Line frequency 60 Hz in USA OTHER SPECIFICATIONS Cal Probe Compensation Voltage 2 V p p 3 square wave 1 kHz nominal CH 2 Y Output Models 2125C 2160C Output Voltage 50 mV div nominal into 50 oh...

Page 9: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 10: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 11: ...IV Control Provides step selection of sweep rate for delayed sweep time base This control has 23 steps from 0 1 S div to 2 S div in a 1 2 5 sequence 25 2125C 2160C DELAY TIME POSition Control Sets sta...

Page 12: ...automatically reverts to triggered sweep operation when an adequate trigger signal is present On the Model 2125C 2160C automatic triggering is applicable to both the main sweep and delayed sweep NORM...

Page 13: ...waveform The other side of the receptacle is the ac return and no waveform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never...

Page 14: ...he red POWER pushbutton Model 2120C 2160C or rotate the POWER control clockwise away from OFF Model 2125C 2160C 3 A trace should appear on the CRT Adjust the trace brightness with the INTENSITY contro...

Page 15: ...iggering TRIGGERING These Oscilloscopes provide versatility in sync triggering for ability to obtain a stable jitter free display in single trace or dual trace operation The proper settings depend upo...

Page 16: ...ontrol is centered the threshold level is set at the approximate average of the signal used as the triggering source Proper adjustment of this control usually synchro nizes the display The TRIG LEVEL...

Page 17: ...ontrols 4 Adjust the amount of horizontal X axis deflection with the CH 1 VOLTS DIV and VARIABLE con trols VIDEO SIGNAL OBSERVATION Setting the COUPLING switch to the TV H or TV V position permits sel...

Page 18: ...a horizontal line Therefore the higher the resistance the closer to horizontal the trace will be Values from 10 to about 5 k are within measurement range Values below 10 will appear to be a dead short...

Page 19: ...OPERATING INSTRUCTIONS Fig 6 Typical Resistive Signatures 17 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 20: ...nductance signatures To test an inductor insert one of the inductor s leads into the white COMP TEST jack and the other into the GND jack make sure that the leads touch the metal walls inside the jack...

Page 21: ...of test leads can be used to connect the COMP TEST and GND jacks to the component s Combinations of Components Using the component test feature it is also possible to observe the signatures of combin...

Page 22: ...installing fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To select the desired line voltage simply insert...

Page 23: ...curacy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is required for calibration adjustments The CAL sig...

Page 24: ...dels 2125C and 2120C this sweep rate is 10 ns div A rise time of less than about four divisions at this sweep speed should be calculated Calculated Measurements For observed rise times of less than 36...

Page 25: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 26: ...play Flashing red dote in top left corner of the first digit is Gate indicator It is light up every time during frequency counter is update NOTE If input signal is not synchronized correctly on CRT di...

Page 27: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 28: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 29: ...try Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes www GlobalTestSu...

Page 30: ...201 4 B K P recision Corp v072914 Printed in Taiwan www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Reviews: