BK Precision 2120C Instruction Manual Download Page 11

PULL INVert:

When pushed in, the polarity of the channel 2 signal
is normal. When pulled out, the polarity of the
channel 2 signal  is  reversed,  thus  inverting the
waveform.

19. CH2 VOLTS/DIV Control.

Vertical attenuator for

channel 2. Provides step adjustment of vertical sensi-
tivity. When channel 2

VARiable

control is set to

CAL

, vertical sensitivity is calibrated in 10 steps from

5 mV/div to 5 V/div in a 1-2-5 sequence. When the
X-Y mode of operation is selected, this control pro-
vides step adjustment of Y-axis sensitivity.

20. CH2 VARiable/PULL X5 MAG Control:

VARiable:

Rotation provides vernier adjustment of channel 2
vertical sensitivity. In the fully-clockwise (

CAL

)

position, the vertical attenuator is calibrated. Coun-
terclockwise rotation decreases gain sensitivity. In
X-Y operation, this control becomes the vernier
Y-axis sensitivity control.

PULL X5 MAG:

When pulled out, increases vertical sensitivity by a
factor of five. Effectively provides two extra sensi-
tivity settings: 2 mV/div and 1 mV/div. In X-Y
mode, increases Y-sensitivity by a factor of five.

21. CH2 (Y) Input Jack.

Vertical input for channel 2.

Y-axis input for X-Y operation.

22.    CH2 AC-GND-DC Switch.

Three-position lever

switch with the following positions:

AC:

Channel 2 input signal is capacitively coupled; dc
component is blocked.

GND:

Opens signal path and grounds input to vertical
amplifier. This provides a zero-volt base line, the
position of which can be used as a reference when
performing dc measurements.

DC:

Direct coupling of channel 2 input signal; both ac
and dc components of signal produce vertical de-
flection.

HORIZONTAL CONTROLS

23. Main Time Base TIME/DIV Control.

Provides step

selection of sweep rate for the main time base. When
the

VARiable Sweep

control is set to

CAL

, sweep rate

is calibrated. This control has 23 steps, from 0.1

µ

S/div

to 2 S/div, in a 1-2-5 sequence.

24.

2125C & 2160C.

DELAY Time Base TIME/DIV

Cont

rol.

Provides step selection of sweep rate for

delayed s

weep time base. This control has 23 steps,

from 0.1 µS

/div to 2 S/div, in a 1-2-5 sequence.

25.

2125

C & 2160C

.

DELAY TIME POSition Control.

Sets s

tarting point of delayed sweep. Clockwis

e

rotation 

causes delayed sweep to begin earlier.

26. VARiable Sweep Control.

Rotation of control is ver-

nier adjustment for sweep rate. In fully clockwise
(

CAL

) position, sweep rate is  calibrated. On the

Model 2125

C

, this control is the vernier adjustment

for both the main and delayed time bases.

27.

POSition/PULL X10 MAG Control.

POSition:

Horizontal (X) position control.

PULL X10 MAG:

Selects ten times sweep magnification when pulled
out, normal when pushed in. Increases maximum
sweep rate to 10 nS/div.

28.

2125

C & 2160C

.

Sweep Mode Switch.

Select

s

sweep (

horizontal) mode. Four-position rotary switch

with t

he following positions:

MAIN:

Only the main sweep operates, with the delayed
sweep inactive.

MIX:

The main and delayed sweep share a single trace;
main sweep occupies the left portion of the display;
delayed sweep occupies the right portion of the
display. The

DELAY TIME POSition

control de-

termines the percentage of display that is main
sweep and the percentage of display that is delayed
sweep (main sweep is usually brighter than the
delayed sweep). Delayed sweep speed cannot be
slower than main sweep speed.

DELAY:

Only delayed sweep operates, while main sweep
stays inactive.

DELAY TIME POSition

control

determines the starting point of the delayed sweep.

X-Y:

Used with the

VERTical MODE

switch and

Trig-

ger SOURCE

switch to select X-Y operating

mode. The channel 1 input becomes the X-axis and
the channel 2 input becomes the Y-axis. Trigger
source and coupling are disabled in this mode.

29.

2120

C

Only.

X-Y Switch.

Used with the

VERTical

MODE

switch and

Trigger SOURCE

switch to se-

lect X-Y operating mode. The channel 1 input be-
comes the X-axis and the channel 2 input becomes the
Y-axis. Trigger source and coupling are disabled in
this mode.

CONTROLS AND INDICATORS

9

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for 2120C

Page 1: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return path 5 When using a probe touch only the...

Page 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 4: ...ring 13 Magnified Sweep Operation 15 OPERATING INSTRUCTIONS Continued X Y Operation 15 Video Signal Observation 15 Application Guidebook 15 Delayed Sweep Operation 15 Component Test Operation 16 MAINT...

Page 5: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 6: ...Y Operation Channel 1 can be applied as horizontal deflection X axis while channel 2 provides vertical deflection Y axis Built in Probe Adjust Square Wave A 2 V p p 1 kHz square wave generator permit...

Page 7: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 8: ...equency Line frequency 60 Hz in USA OTHER SPECIFICATIONS Cal Probe Compensation Voltage 2 V p p 3 square wave 1 kHz nominal CH 2 Y Output Models 2125C 2160C Output Voltage 50 mV div nominal into 50 oh...

Page 9: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 10: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 11: ...IV Control Provides step selection of sweep rate for delayed sweep time base This control has 23 steps from 0 1 S div to 2 S div in a 1 2 5 sequence 25 2125C 2160C DELAY TIME POSition Control Sets sta...

Page 12: ...automatically reverts to triggered sweep operation when an adequate trigger signal is present On the Model 2125C 2160C automatic triggering is applicable to both the main sweep and delayed sweep NORM...

Page 13: ...waveform The other side of the receptacle is the ac return and no waveform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never...

Page 14: ...he red POWER pushbutton Model 2120C 2160C or rotate the POWER control clockwise away from OFF Model 2125C 2160C 3 A trace should appear on the CRT Adjust the trace brightness with the INTENSITY contro...

Page 15: ...iggering TRIGGERING These Oscilloscopes provide versatility in sync triggering for ability to obtain a stable jitter free display in single trace or dual trace operation The proper settings depend upo...

Page 16: ...ontrol is centered the threshold level is set at the approximate average of the signal used as the triggering source Proper adjustment of this control usually synchro nizes the display The TRIG LEVEL...

Page 17: ...ontrols 4 Adjust the amount of horizontal X axis deflection with the CH 1 VOLTS DIV and VARIABLE con trols VIDEO SIGNAL OBSERVATION Setting the COUPLING switch to the TV H or TV V position permits sel...

Page 18: ...a horizontal line Therefore the higher the resistance the closer to horizontal the trace will be Values from 10 to about 5 k are within measurement range Values below 10 will appear to be a dead short...

Page 19: ...OPERATING INSTRUCTIONS Fig 6 Typical Resistive Signatures 17 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 20: ...nductance signatures To test an inductor insert one of the inductor s leads into the white COMP TEST jack and the other into the GND jack make sure that the leads touch the metal walls inside the jack...

Page 21: ...of test leads can be used to connect the COMP TEST and GND jacks to the component s Combinations of Components Using the component test feature it is also possible to observe the signatures of combin...

Page 22: ...installing fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To select the desired line voltage simply insert...

Page 23: ...curacy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is required for calibration adjustments The CAL sig...

Page 24: ...dels 2125C and 2120C this sweep rate is 10 ns div A rise time of less than about four divisions at this sweep speed should be calculated Calculated Measurements For observed rise times of less than 36...

Page 25: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 26: ...play Flashing red dote in top left corner of the first digit is Gate indicator It is light up every time during frequency counter is update NOTE If input signal is not synchronized correctly on CRT di...

Page 27: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 28: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 29: ...try Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes www GlobalTestSu...

Page 30: ...201 4 B K P recision Corp v072914 Printed in Taiwan www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Reviews: