BK Precision 2120C Instruction Manual Download Page 14

OPERATING TIPS

The following recommendations will help obtain the best

performance from the oscilloscope.

1. Always use the probe ground clips for best results,

attached to a circuit ground point near the point of
measurement. Do not rely solely on an external ground
wire in lieu of the probe ground clips as undesired
signals may be introduced.

2. Avoid the following operating conditions:

a. Direct sunlight.

b. High temperature and humidity.

c. Mechanical vibration.

d. Electrical noise and strong magnetic fields, such as

near large motors, power supplies, transformers,
etc.

3. Occasionally check trace rotation, probe compensa-

tion, and calibration accuracy of the oscilloscope using
the procedures found in the MAINTENANCE section
of this manual.

4. Terminate the output of a signal generator into its

characteristic impedance to minimize ringing, espe-
cially if the signal has fast edges such as square waves
or pulses. For example, the typical 50

output of a

square wave generator should be terminated into an
external 50

terminating load and connected to the

oscilloscope with 50

coaxial cable.

5. Probe compensation adjustment matches the probe to

the input of the scope. For best results, compensation
should be adjusted initially, then the same probe al-
ways used with the same channel. Probe compensation
should be readjusted when a probe from a different
oscilloscope is used.

INITIAL STARTING PROCEDURE

Until you familiarize yourself with the use of all controls,

the settings given here can be used as a reference point to
obtain a trace on the CRT in preparation for waveform
observation.

1. Set these controls as follows:

On both models:

VERTical MODE

to

CH1

.

CH1 AC/GND/DC

to

GND.

Trigger COUPLING

to

AUTO.

Trigger SOURCE

to

CH1.

All

POSition

controls and

INTENSITY

control cen-

tered (pointers facing up).

Main Time Base

control to

1 mS/div.

On the Model 2125

C & 2160C

:

Sweep Mode

switch to

MAIN

.

2. Press the red

POWER

pushbutton (Model 2120

C &

2160C), or r

otate the

POWER

control clockwis

e

away from "

OFF

"

(Model 2125

C & 2160C

).

3. A trace should appear on the CRT. Adjust the trace

brightness  with the

INTENSITY

control, and  the

trace sharpness with the

FOCUS

control.

NOTE

On the  Model 2125

C & 2160C

 you  can use

the

 

BEAM FINDER

pushbutton to locate a

trace that has been moved off the screen by
the

POSition

controls. When the button is

pushed, a compressed version of the trace
is brought into view which indicates the
location of the trace.

SINGLE TRACE DISPLAY

Either channel 1 or channel 2 may be used for single-trace

operation. To observe a waveform on channel 1:

1. Perform the steps of the “Initial Starting Procedure”.

2. Connect the probe to the

CH 1 (X)

input jack.

3. Connect the probe ground clip to the chassis or com-

mon of the equipment under test. Connect the probe
tip to the point of measurement.

4. Move the

CH1 AC/GND/DC

switch out of the

GND

position to either

DC

or

AC

.

5. If no waveforms appear, increase the sensitivity by

turning the

CH 1 VOLTS/DIV

control clockwise to a

position that gives 2 to 6 divisions vertical deflection.

6. Position the waveform vertically as desired using the

CH1 POSition

control.

7. The display on the CRT may be unsynchronized. Refer

to the “Triggering” paragraphs in this section for pro-
cedures on setting triggering and sweep time controls
to obtain a stable display showing the desired number
of waveforms.

DUAL TRACE DISPLAY

In observing simultaneous waveforms on channel 1 and

2, the waveforms are usually related in frequency, or one of
the waveforms is synchronized to the other, although the
basic frequencies are different. To observe two such related
waveforms simultaneously, perform the following:

1. Connect probes to both the

CH 1 (X)

and

CH 2 (Y)

input jacks.

2. Connect the ground clips of the probes to the chassis

or common of the equipment under test. Connect the
tips of the probes to the two points in the circuit where
waveforms are to be measured.

OPERATING INSTRUCTIONS

12

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for 2120C

Page 1: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return path 5 When using a probe touch only the...

Page 3: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 4: ...ring 13 Magnified Sweep Operation 15 OPERATING INSTRUCTIONS Continued X Y Operation 15 Video Signal Observation 15 Application Guidebook 15 Delayed Sweep Operation 15 Component Test Operation 16 MAINT...

Page 5: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 6: ...Y Operation Channel 1 can be applied as horizontal deflection X axis while channel 2 provides vertical deflection Y axis Built in Probe Adjust Square Wave A 2 V p p 1 kHz square wave generator permit...

Page 7: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 8: ...equency Line frequency 60 Hz in USA OTHER SPECIFICATIONS Cal Probe Compensation Voltage 2 V p p 3 square wave 1 kHz nominal CH 2 Y Output Models 2125C 2160C Output Voltage 50 mV div nominal into 50 oh...

Page 9: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 10: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 11: ...IV Control Provides step selection of sweep rate for delayed sweep time base This control has 23 steps from 0 1 S div to 2 S div in a 1 2 5 sequence 25 2125C 2160C DELAY TIME POSition Control Sets sta...

Page 12: ...automatically reverts to triggered sweep operation when an adequate trigger signal is present On the Model 2125C 2160C automatic triggering is applicable to both the main sweep and delayed sweep NORM...

Page 13: ...waveform The other side of the receptacle is the ac return and no waveform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never...

Page 14: ...he red POWER pushbutton Model 2120C 2160C or rotate the POWER control clockwise away from OFF Model 2125C 2160C 3 A trace should appear on the CRT Adjust the trace brightness with the INTENSITY contro...

Page 15: ...iggering TRIGGERING These Oscilloscopes provide versatility in sync triggering for ability to obtain a stable jitter free display in single trace or dual trace operation The proper settings depend upo...

Page 16: ...ontrol is centered the threshold level is set at the approximate average of the signal used as the triggering source Proper adjustment of this control usually synchro nizes the display The TRIG LEVEL...

Page 17: ...ontrols 4 Adjust the amount of horizontal X axis deflection with the CH 1 VOLTS DIV and VARIABLE con trols VIDEO SIGNAL OBSERVATION Setting the COUPLING switch to the TV H or TV V position permits sel...

Page 18: ...a horizontal line Therefore the higher the resistance the closer to horizontal the trace will be Values from 10 to about 5 k are within measurement range Values below 10 will appear to be a dead short...

Page 19: ...OPERATING INSTRUCTIONS Fig 6 Typical Resistive Signatures 17 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 20: ...nductance signatures To test an inductor insert one of the inductor s leads into the white COMP TEST jack and the other into the GND jack make sure that the leads touch the metal walls inside the jack...

Page 21: ...of test leads can be used to connect the COMP TEST and GND jacks to the component s Combinations of Components Using the component test feature it is also possible to observe the signatures of combin...

Page 22: ...installing fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To select the desired line voltage simply insert...

Page 23: ...curacy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is required for calibration adjustments The CAL sig...

Page 24: ...dels 2125C and 2120C this sweep rate is 10 ns div A rise time of less than about four divisions at this sweep speed should be calculated Calculated Measurements For observed rise times of less than 36...

Page 25: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 26: ...play Flashing red dote in top left corner of the first digit is Gate indicator It is light up every time during frequency counter is update NOTE If input signal is not synchronized correctly on CRT di...

Page 27: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 28: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 29: ...try Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes www GlobalTestSu...

Page 30: ...201 4 B K P recision Corp v072914 Printed in Taiwan www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Reviews: