36
MAINTENANCE (CONT.)
3.
Connect the Waveform Monitor and the Vectorscope to the
COMPOSITE VIDEO
jack (terminated into 75 S2). Set the
Waveform Monitor response to FLAT and the sweep to 2H.
Set the variable amplitude control on the Waveform Monitor
to the CAL position. Set the sync to internal, and turn on the
DC restorer.
Color Subcarrier Adjustment
1.
Connect the frequency counter at the junction of R95 and pin
6 of IC1 (the right hand side of R95) and adjust C4 for a
reading of 3.579545MHz ±3Hz on the frequency counter.
Sound Subcarrier Adjustment
1.
Press the 4.5MHz switch on the Generator
2.
Connect the frequency counter at the junction of R81 and pin
13 of IC18(the back lead of R81) and adjust L4 for a reading
of 4.49MHz ±10kHz on the frequency counter.
3.
Release the
4.5 MHz
switch on the NTSC Generator.
I-F and RF Adjustments
1.
Make sure that the
CH 4/CH 3
switch is disengaged (in the
CH
3
position), connect the frequency counter to the
IF/RF
jack, and adjust the core of L2 fully clockwise (to the
bottom) or until the frequency counter reads below
61.245MHz.
2.
Now adjust L2 counterclockwise until the frequency
counter reads 61.250MHz ±5KHz. Turn the core an
additional 1/2 turn counter-clockwise. The reading on the
frequency counter should retrain stable.
3.
Press the
CH 4/CH 3
switch on the NTSC Generator (so that it
is in the
CH 4
position).
4.
With the frequency counter still connected to the
IF/RF
jack,
adjust the core of L3 fully clockwise (to the bottom) or until the
frequency counter reads below 67.245MHz.
5.
Now adjust L3 counterclockwise until the frequency counter
reads 67.250MHz ±5KHz. Turn the core an additional 1/2 turn
counter-clockwise. The reading on the frequency counter
should remain stable.
6.
Press the
IF/RF
switch on the Generator (so that it is in the
IF
position).
7.
Adjust the core of LI fully clockwise (to the bottom) or until
the frequency counter reads below 45.745MHz.
8.
Now adjust L I counterclockwise
'
until the frequency counter
reads 45.750MHz ±5 KHz. Turn the core an additional 1/2 turn
counter-clockwise. The reading on the frequency counter
should remain stable.
9.
Release the
IF/RF
switch (return to channel 4 rf operation).
The counter should now read 67.250MHz ±5KHz. If the
reading is unstable, repeat step 5.
10.
Release the
CH 4/CH 3
switch. The switch should now be in
the
CH 3
position. The frequency counter should now read
61.250MHz ±5KHz. If the reading is unstable, repeat step 2.
Sync Pulse Adjustment
1.
Set the
COMPOSITE VIDEO LEVEL
control to
CAL
position and adjust VR8 for sync pulse amplitude of -40 IRE on
the Waveform Monitor (baseline on zero line).
Summary of Contents for 1249B
Page 1: ...INSTRUCTION MANUAL Model 1249B NTSC GENERATOR ...
Page 16: ...16 THE NTSC COLOR VIDEO SIGNAL CONT Fig 3 Elements of Color Television Signal ...
Page 20: ...20 CONTROLS AND INDICATORS CONT Fig 4 Controls and Indicators ...
Page 35: ...35 MAINTENANCE CONT R7 Fig 8 Adjustment and Test Point Locations ...