background image

 3

8817 • Disposal

Connecting the Transducer

The transducer is connected to the scanner using the array Transducer Socket on the 
scanner. To connect, the transducer plug’s locking lever should first be in a horizontal 
position. Align the plug to the scanner socket and insert securely. Turn the locking lever 
clockwise to lock in place.

When connected the transducer complies with Type BF requirements of EN60601-1 (IEC 
60601-1).

Changing Frequency

The Multi-Frequency Imaging (MFI) facility enables you to select the scanning frequency. 
See the applicable scanner user guide for instructions. The selected frequency is displayed 
at the top of the screen.

Using a Transducer Cover

The transducer should be enclosed in a transducer cover or a standard condom. See the 
Product Data sheet for a list of available transducer covers.

Apply gel to the tip of the transducer. This improves the screen images by preventing image 
artifacts caused by air bubbles.

Pull the transducer cover over the transducer.

Gel also creates a good acoustic contact between the skin and the transducer; therefore, 
apply a small amount to the outside of the cover prior to scanning. Re-apply the gel 
frequently to ensure good screen images.

Changing Orientation

To change the orientation of the image on the monitor, refer to the applicable scanner user 
guide for instructions.

Disposal

When the transducer is scrapped at the end of its life, national rules for the relevant 
material in each individual land must be followed. Within the EU, when you discard the 

WARNING

Keep all plugs and sockets absolutely dry at all times.

WARNING

Because of reports of severe allergic reactions to medical devices containing latex (natural 
rubber), FDA is advising health-care professionals to identify their latex-sensitive patients 
and be prepared to treat allergic reactions promptly. 

WARNING

Use only water-soluble agents or gels. Petroleum or mineral oil-based materials may harm 
the cover material.