
E14C4t User Guide
(16-01265-EN-07)
Imaging with E14C4t
9
Apply sterile gel to the tip of the transducer or fill the cover with 1 to 2 ml of sterile
water. This improves screen imaging by preventing image artifacts caused by air
bubbles.
Gel also creates a good acoustic contact between the skin and the transducer;
therefore, apply a small amount to the outside of the cover prior to imaging and re-
apply frequently.
Follow these precautions when putting sterile covers on a transducer:
•
Wear sterile gloves.
•
When using a puncture attachment, place it gently over the cover and secure it,
following the instructions for the puncture attachment.
•
Verify that the cover has not been damaged in the process. If it has, repeat the
procedure with a new transducer cover.
Using the Transducer Control Buttons
To change the imaging plane, press the button corresponding to that plane (see Figure
1). Pressing the button activates (starts) or freezes (stops) imaging in that plane. One
button enables the sagittal or endfire array. A long press on this button also activates
endfire imaging.
A second button enables the transverse array.
Changing Orientation
To change the orientation of the image on the monitor, refer to the applicable system
user guide for instructions.
Imaging with E14C4t
Simultaneous Biplane Imaging
E14C4t can transmit transverse (T) and sagittal (S) images simultaneously. When
you press bk3000/bk5000’s
Split
key or tap
LIVE DUAL
on bkSpecto, simultaneous
live transmission is automatically activated. On the bk3000/bk5000 screen, this is
indicated by a green dot in front of
Simultan
.
Caution
T-c3
Use only water-based gel (sterile if you are using a sterile transducer cover). Products
containing parabens, petroleum, or mineral oils may harm the transducer or transducer
cover.
WARNING
Colo-w1
Do not use excessive force during insertion. Do not make excessive lateral movements
during or after insertion. Risk of injury or tissue damage to the patient could occur under
certain circumstances. A digital palpation of the rectum may need to be carried out by a
clinician prior to insertion or use of the probe as a precautionary measure.
Summary of Contents for E14C4t
Page 1: ...For Professional Users Only 16 01265 EN 07 June 2019 E14C4t Transducer User Guide...
Page 4: ...4...
Page 19: ......
Page 20: ......